PURELY YOURS
Report
- Report Number
- 3009974348-2015-00139
- Event Type
- Injury
- Date Received
- July 7, 2015
- Date of Event
- May 19, 2014
- Report Date
- May 21, 2014
- Manufacturer
- AMEDA INC.
- Product Code
- HGX
- PMA / PMN Number
- K973501
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
PRODUCT WAS EVALUATED FOR EVIDENCE OF ALLEGATION. THE RETURNED AMEDA PURELY YOURS BREAST PUMP MET AMEDA SPECIFICATIONS FOR BOTH SUCTION AND SPEED, AND PASSED VISUAL INSPECTION STANDARDS. NO EVIDENCE OF MALFUNCTION WAS OBSERVED.
AS PART OF AMEDA, INC.'S QUALITY MANAGEMENT SYSTEM ACTIVITIES, A REVIEW OF HISTORICAL COMPLAINTS WAS CONDUCTED. IT WAS DETERMINED THAT THIS COMPLAINT SHOULD HAVE BEEN REPORTED TO FDA. CUSTOMER CONTACTED AMEDA, INC. TO REPORT UNEVEN, LOWER SUCTION WHILE USING THE PURELY YOURS BREAST PUMP. THIS ISSUE RESULTED IN DECREASED MILK OUTPUT. CUSTOMER USES THIS BREAST PUMP EXCLUSIVELY BUT REPORTS THE PUMP DOES NOT EFFECTIVELY DRAIN HER BREASTS. CUSTOMER WAS DIAGNOSED WITH LEFT BREAST MASTITIS ON (B)(6) 2014 BY HER HEALTH CARE PROVIDER AND PRESCRIBED ORAL ANTIBIOTICS TO RESOLVE INFECTION. CUSTOMER STATES THE INFECTION IS RESOLVED WITHIN 2 DAYS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 440854 | PURELY YOURS | ELECTRID BREAST PUMP | HGX | AMEDA INC. | 24502082 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |