FDA Adverse Event Injury Summary report: N

PURELY YOURS

MDR report key: 4921994 · Received July 7, 2015

Report

Report Number
3009974348-2015-00139
Event Type
Injury
Date Received
July 7, 2015
Date of Event
May 19, 2014
Report Date
May 21, 2014
Manufacturer
AMEDA INC.
Product Code
HGX
PMA / PMN Number
K973501
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT WAS EVALUATED FOR EVIDENCE OF ALLEGATION. THE RETURNED AMEDA PURELY YOURS BREAST PUMP MET AMEDA SPECIFICATIONS FOR BOTH SUCTION AND SPEED, AND PASSED VISUAL INSPECTION STANDARDS. NO EVIDENCE OF MALFUNCTION WAS OBSERVED.

Description of Event or Problem · 1

AS PART OF AMEDA, INC.'S QUALITY MANAGEMENT SYSTEM ACTIVITIES, A REVIEW OF HISTORICAL COMPLAINTS WAS CONDUCTED. IT WAS DETERMINED THAT THIS COMPLAINT SHOULD HAVE BEEN REPORTED TO FDA. CUSTOMER CONTACTED AMEDA, INC. TO REPORT UNEVEN, LOWER SUCTION WHILE USING THE PURELY YOURS BREAST PUMP. THIS ISSUE RESULTED IN DECREASED MILK OUTPUT. CUSTOMER USES THIS BREAST PUMP EXCLUSIVELY BUT REPORTS THE PUMP DOES NOT EFFECTIVELY DRAIN HER BREASTS. CUSTOMER WAS DIAGNOSED WITH LEFT BREAST MASTITIS ON (B)(6) 2014 BY HER HEALTH CARE PROVIDER AND PRESCRIBED ORAL ANTIBIOTICS TO RESOLVE INFECTION. CUSTOMER STATES THE INFECTION IS RESOLVED WITHIN 2 DAYS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
440854 PURELY YOURS ELECTRID BREAST PUMP HGX AMEDA INC. 24502082

Patients

Seq Age Sex Outcome Treatment
1 Other