SPECTRUM INFUSION PUMP
Report
- Report Number
- 1314492-2015-07700
- Event Type
- Malfunction
- Date Received
- July 17, 2015
- Date of Event
- June 1, 2015
- Report Date
- June 25, 2015
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Removal / Correction Number
- Z-1484-2014
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE DEVICE WAS FOUND OUT OF SPECIFICATION IN RELATION TO THE REPORTED SYSTEM ERROR 322, WHICH WAS NOT REPRODUCED. SYSTEM ERROR 322 WAS CONFIRMED THROUGH A REVIEW OF THE EVENT HISTORY LOG. EVALUATION FOUND THE CAUSED THE LOWER HOOK GAP TO BE OUT OF SPECIFICATION. THE LOWER HOOK GAP WAS ADJUSTED. THE SOFTWARE WAS ALSO UPGRADED TO CORRECT THIS ISSUE PER THE APPROVED REMEDIAL ACTION ACTIVITIES. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE. BAXTER HAS INITIATED A CAPA TO FURTHER INVESTIGATE THIS ISSUE. ADDITIONAL INFORMATION: CALIBRATION ISSUE.
IT WAS REPORTED THAT A SPECTRUM PUMP DISPLAYED SYSTEM ERROR 322. IT WAS ALSO REPORTED THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 466696 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORPORATION | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |