FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 4920751
·
Received July 17, 2015
Report
- Report Number
- 0001811755-2015-02598
- Event Type
- Malfunction
- Date Received
- July 17, 2015
- Date of Event
- June 23, 2015
- Report Date
- June 24, 2015
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K112593
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE FAILURE FOR HEAT WAS CONFIRMED THROUGH FUNCTIONAL EVALUATION, DISASSEMBLY AND VISUAL INSPECTION. THROUGH VISUAL INSPECTION, THE SET SCREW MOVE.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY, THE CORE IMPACTION DRILL WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES REPORTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE AT THE USER FACILITY, THE CORE IMPACTION DRILL WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 465615 | CORE IMPACTION DRILL | DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE | ERL | STRYKER INSTRUMENTS-KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |