FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 4920480 · Received July 17, 2015

Report

Report Number
2939301-2015-29770
Event Type
Malfunction
Date Received
July 17, 2015
Report Date
July 7, 2015
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1. THE LAY USER/PATIENTS TEST STRIPS HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE TEST STRIPS PASSED ALL TESTING WITH NO FAULTS FOUND. THE REPORTED ISSUE COULD NOT BE CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING OBTAINING AN INACCURATE LOW CONTROL SOLUTION RESULT. THE REPORTER CLAIMED OBTAINING A CONTROL RESULT OF ¿100 MG/DL¿ WHICH FELL BELOW THE SPECIFIED CONTROL SOLUTION RANGE PRINTED ON THE TEST STRIP VIAL. THIS COMPLAINT IS BEING REPORTED BECAUSE THE ALLEGED INACCURACY CONTROL ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
464623 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3679296

Patients

Seq Age Sex Outcome Treatment
1