FDA Adverse Event Injury Summary report: N

PANTA PARTIALLY THREADED 5MM SCREW - 25MM LENGTH

MDR report key: 4918006 · Received July 13, 2015

Report

Report Number
9615741-2015-00035
Event Type
Injury
Date Received
July 13, 2015
Date of Event
May 30, 2015
Report Date
June 15, 2015
Manufacturer
NEWDEAL SAS
Product Code
HSB
PMA / PMN Number
K091788
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IT WAS REPORTED THE PANTA SCREWS BROKE AND THE PATIENT WOULD NEED REVISION SURGERY DUE TO THE DEVICES BREAKING. THERE ARE 2 PRODUCTS, 4 LOTS ASSOCIATED WITH THIS EVENT. TWO ADDITIONAL REPORTS WILL BE SUBMITTED TO CAPTURE THE 2 ADDITIONAL LOTS. THIS IS REPORT 1 OF 4 SUBMISSIONS FOR THIS EVENT. ADDITIONAL INFORMATION RECEIVED 21SEP2015: REVISION SURGERY OCCURRED SEPT, 2015. THE BROKEN SCREWS WERE REMOVED AND COLLECTED FOR RETURN. PRODUCT ID UPDATED.

Additional Manufacturer Narrative · 1

INTEGRA HAS COMPLETED THEIR INTERNAL INVESTIGATION ON 6AUG2015. THE ADDITIONAL INVESTIGATION ACTIVITIES INCLUDED: METHODS: -REVIEW OF DEVICE HISTORY RECORDS. -REVIEW OF COMPLAINT HISTORY. RESULTS: A REVIEW OF THE DESIGN SPECIFICATIONS AND DESIGN CHANGES WAS PERFORMED. NO SIGNIFICANT DESIGN CHANGE WAS DONE ON DIMENSIONS OR RAW MATERIAL ON SPECIFICATIONS DURING THE LAST TWO YEARS. A REVIEW OF THE DEVICE HISTORY CANNOT BE PERFORMED AS NO LOT NUMBER IS PROVIDED TO NEWDEAL. A REVIEW OF THE COMPLAINT SYSTEM WAS PERFORMED. THIS IS THE FIRST INCIDENT (TWO SCREWS ARE CONCERNED) REPORTED TO NEWDEAL ABOUT A BREAKAGE OF A PANTA® NAIL PARTIALLY THREADED SCREW DIAMETER 5MM FOR THE PAST TWO YEARS. DURING THE SAME TIME PERIOD, ABOUT (B)(4) PANTA® NAIL PARTIALLY THREADED SCREWS DIAMETER 5MM WERE SOLD. THE COMPLAINT RATE FOR THIS KIND OF INCIDENT DURING THE STATED TIME PERIOD IS (B)(4) PERCENT. LOT FAILURE RATE CANNOT BE CALCULATED AS NO LOT NUMBER IS PROVIDED. CONCLUSION: GIVEN THE DESCRIPTION OF THE EVENT AND THE OBSERVATIONS MADE DURING THE INVESTIGATION, THE ROOT CAUSE OF THE INCIDENT CANNOT BE DETERMINED AS NO LOT NUMBER WAS PROVIDED AND NO PRODUCT WAS RETURNED FOR ANALYSIS. NO ANOMALY WAS FOUND DURING DOCUMENTARY INVESTIGATION.

Additional Manufacturer Narrative · 1

THE DEVICE INVOLVED IN THE REPORTED INCIDENT IS NOT EXPECTED TO BE RECEIVED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED UPON THE REPORTED INFORMATION.

Additional Manufacturer Narrative · 1

INTEGRA HAS COMPLETED THEIR INTERNAL INVESTIGATION ON 12/04/2015. THE INVESTIGATION INCLUDED: METHODS: EVALUATION OF ACTUAL DEVICE, REVIEW OF DEVICE HISTORY RECORDS, REVIEW OF COMPLAINTS HISTORY. RESULTS: PRODUCTS WERE RETURNED FOR EVALUATION. A VISUAL INSPECTION WAS PERFORMED. ANOMALIES WERE OBSERVED. SCREWS P/N 511065S & 511070S WERE BREAKAGE AT THE SHAFT DIAMETER. NO ANOMALY WAS OBSERVED FOR P/N 511025S. A MEASUREMENT INSPECTION WAS PERFORMED. NO ANOMALY WAS OBSERVED. SHAFT DIAMETER Ø5MM WAS MEASURED FOR EACH BROKEN SCREWS. FOR P/N 511065S THE SHAFT DIAMETER WAS MEASURED AT 4,90MM. FOR P/N 511070S THE SHAFT DIAMETER WAS MEASURED AT 4,92MM. BOTH DIAMETERS ARE IN COMPLIANCE WITH NEWDEAL SPECIFICATIONS Ø5 (0/-0, 15 MM). NO ANOMALY WAS OBSERVED FOR THE PANTA NAIL 500180S. A REVIEW OF THE DESIGN SPECIFICATIONS AND DESIGN CHANGES WAS PERFORMED. TECHNICAL SPECIFICATIONS ARE CLEARLY ESTABLISHED IN TERM OF EXTERNAL DIAMETER, THREADED DIAMETER, BODY DIAMETER, AND RAW MATERIAL AND ASSOCIATED ISO STANDARD. NO SIGNIFICANT DESIGN CHANGE WAS DONE ON DIMENSIONS OR RAW MATERIAL ON SPECIFICATIONS FOR PANTA® NAIL PARTIALLY THREADED SCREW DIAMETER 5MM DURING THE LAST TWO YEARS. A REVIEW OF THE DEVICE HISTORY RECORD IS PERFORMED FOR P/N 511065S. MANUFACTURING LOT NUMBER F3HB WAS MANUFACTURED IN APRIL 2012. NO ANOMALY WAS DETECTED REGARDING THE INCIDENT. A REVIEW OF THE DEVICE HISTORY RECORD IS PERFORMED FOR P/N 511025S. MANUFACTURING LOT NUMBER F4NK WAS MANUFACTURED IN FEBRUARY 2013. NO ANOMALY WAS DETECTED REGARDING THE INCIDENT. A REVIEW OF THE DEVICE HISTORY CANNOT BE PERFORMED FOR P/N511070S AS NO LOT NUMBER IS PROVIDED. A REVIEW OF THE DEVICE HISTORY RECORD IS PERFORMED FOR P/N 500180S. MANUFACTURING LOT NUMBER F33C WAS MANUFACTURED IN DECEMBER 2012. NO ANOMALY WAS DETECTED REGARDING THE INCIDENT. A REVIEW OF THE COMPLAINT SYSTEM WAS PERFORMED. THIS IS THE FIRST INCIDENT (THREE SCREWS ARE CONCERNED AND A PANTA NAIL) REPORTED TO NEWDEAL ABOUT A BREAKAGE OF A PANTA® NAIL PARTIALLY THREADED SCREW DIAMETER 5MM FOR THE PAST TWO YEARS. DURING THE SAME TIME PERIOD, ABOUT 5604 PANTA® NAIL PARTIALLY THREADED SCREWS DIAMETER 5MM WERE SOLD. THE COMPLAINT RATE FOR THIS KIND OF INCIDENT DURING THE STATED TIME PERIOD IS 0.0398 PERCENT. CONCLUSION: GIVEN THE DESCRIPTION OF THE EVENT AND THE OBSERVATIONS MADE DURING THE INVESTIGATION, THE ROOT CAUSE OF THE INCIDENT CANNOT BE DETERMINED. REGARDING THE OBSERVATION MADE DURING THIS INVESTIGATION, THE PANTA NAIL INVOLVED IN THIS CASE IS NOT RELATED TO THE CAUSE OF INCIDENT. NO ANOMALY WAS OBSERVED DURING FAILURE ANALYSIS AND REVIEW OF QUALITY RECORDS.

Description of Event or Problem · 1

IT WAS REPORTED THE PATIENT WILL NEED REVISION SURGERY DUE TO THE DEVICES BREAKING. THIS IS THE FIRST OF TWO RELATED REPORTS. IT WAS REPORTED THE PANTA NAIL SYSTEM WAS IMPLANTED 5 MONTHS PRIOR TO THIS EVENT. THE PATIENT RECENTLY RETURNED TO THE SURGEON WITH SEVERE FOOT PAIN. IT WAS OBSERVED THAT BOTH THE CALCANEAL SCREWS BROKE AND ALSO TALUS BONE COMPRESSION AT UNHEALED SURGICAL SITE. THE DEVICES WERE IMPLANTED ON (B)(6) 2014. THE EVENT WAS OBSERVED ON (B)(6) 2015. A SECOND SURGERY WILL BE PERFORMED (B)(6) 2015. THE PATIENT HAD SYMPTOMS OF SEVERE PAIN AND NON-UNION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
453081 PANTA PARTIALLY THREADED 5MM SCREW - 25MM LENGTH N/A HSB NEWDEAL SAS F3H8

Patients

Seq Age Sex Outcome Treatment
1 29 YR Required Intervention