FDA Adverse Event Malfunction Summary report: N

LOCKING NUT DRIVER, HEXALOBE, SHORT (STAINLESS STEEL)

MDR report key: 4917151 · Received July 15, 2015

Report

Report Number
2027467-2015-00139
Event Type
Malfunction
Date Received
July 15, 2015
Date of Event
October 16, 2013
Report Date
June 12, 2015
Manufacturer
ALPHATEC SPINE INC
Product Code
HXX
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MDR BEING SUBMITTED AS A RESULT OF A RETROSPECTIVE COMPLAINT REVIEW. REVIEW OF THE DEVICE HISTORY RECORDS REVEALED NO MANUFACTURING, PROCESSING OR DESIGN RELATED IRREGULARITIES. THE INSTRUMENT WAS FOUND TO BE PROPERLY MANUFACTURED AND RELEASED IN ACCORDANCE WITH THE DEVICE MASTER RECORD. VISUAL INSPECTION OF THE RETURNED DRIVER FOUND THE DISTAL HEXALOBE TIP HAD FRACTURED AND BECAME DETACHED. BOTH THE RAW MATERIAL AND HARDNESS WERE INSPECTED AND DETERMINED TO MEET PRINT SPECIFICATIONS. RAW MATERIAL WAS TESTED VIA DEFRACTOMETER WHICH CONFIRMED THE MATERIAL TO BE 440SS. THE SUPPLIER CERTIFICATION AND IN-HOUSE HARDNESS TESTING INDICATED THE HRC TO BE A MINIMUM OF 55 AS REQUIRED. ROOT CAUSE APPEARS TO BE THE USE OF THE DRIVER WITH THE INCORRECT TORQUE HANDLE WHICH ALLOW ADDITIONAL FORCE TO BE APPLIED TO THE INSTRUMENT.

Description of Event or Problem · 1

THE TIP OF THE INSTRUMENT BROKE DURING SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
460916 LOCKING NUT DRIVER, HEXALOBE, SHORT (STAINLESS STEEL) HXX HXX ALPHATEC SPINE INC IT-SD1009 21403-1

Patients

Seq Age Sex Outcome Treatment
1 Other