DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2015-05883
- Event Type
- Malfunction
- Date Received
- July 15, 2015
- Date of Event
- June 22, 2015
- Report Date
- June 22, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT TRANSMITTER WAS NOT RETURNED TO DEXCOM. THE RECEIVER (PART NUMBER STR-GL-001/SERIAL NUMBER (B)(4)/LOT NUMBER 5150457), BEING USED WITH THE COMPLAINT TRANSMITTER, WAS RETURNED ON 07/20/2015. THE RETURNED RECEIVER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION; HOWEVER, A REVIEW OF THE DOWNLOADED RECEIVER LOG CONFIRMED THE REPORTED EVENT OF AN INTERMITTENT OUT OF RANGE SIGNAL. A ROOT CAUSE COULD NOT BE DETERMINED.
(B)(4).
PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015 TO REPORT THAT ON (B)(6) 2015, PATIENT EXPERIENCED AN INTERMITTENT OUT OF RANGE SIGNAL. NO ADDITIONAL PATIENT OR EVENT INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 459983 | DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | 9438-05 | 5197294 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |