FDA Adverse Event Malfunction Summary report: N

CONTOUR LINK

MDR report key: 4915398 · Received July 15, 2015

Report

Report Number
1826988-2015-00343
Event Type
Malfunction
Date Received
July 15, 2015
Date of Event
February 2, 2015
Report Date
June 24, 2015
Manufacturer
BAYER HEALTHCARE LLC
Product Code
NBW
PMA / PMN Number
K110587
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IN SOME COUNTRIES OUTSIDE THE US, CUSTOMER INFORMATION IS NOT PROVIDED DUE TO PRIVACY LAWS.

Additional Manufacturer Narrative · 1

THE RETURNED METER WAS EVALUATED AND CONFIRMED FOR THE PROBLEM OF STORING RESULTS TO THE MEMORY DUE TO A LOW BATTERY CONDITION. ONCE THE BATTERIES WERE REPLACED RESULTS WERE STORED IN THE MEMORY.

Description of Event or Problem · 1

(B)(6) CUSTOMER STATED SOME OF HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR LINK. NO ADVERSE EVENT ALLEGED. THE CUSTOMER WAS ASKED TO RETURN THE METER FOR INVESTIGATION. A REPLACEMENT METER WAS SENT TO HIM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
460998 CONTOUR LINK BLOOD GLUCOSE METER NBW BAYER HEALTHCARE LLC 7778

Patients

Seq Age Sex Outcome Treatment
1