FDA Adverse Event
Malfunction
Summary report: N
CONTOUR LINK
MDR report key: 4915398
·
Received July 15, 2015
Report
- Report Number
- 1826988-2015-00343
- Event Type
- Malfunction
- Date Received
- July 15, 2015
- Date of Event
- February 2, 2015
- Report Date
- June 24, 2015
- Manufacturer
- BAYER HEALTHCARE LLC
- Product Code
- NBW
- PMA / PMN Number
- K110587
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
IN SOME COUNTRIES OUTSIDE THE US, CUSTOMER INFORMATION IS NOT PROVIDED DUE TO PRIVACY LAWS.
Additional Manufacturer Narrative · 1
THE RETURNED METER WAS EVALUATED AND CONFIRMED FOR THE PROBLEM OF STORING RESULTS TO THE MEMORY DUE TO A LOW BATTERY CONDITION. ONCE THE BATTERIES WERE REPLACED RESULTS WERE STORED IN THE MEMORY.
Description of Event or Problem · 1
(B)(6) CUSTOMER STATED SOME OF HIS BLOOD GLUCOSE READINGS WERE NOT BEING STORED IN THE MEMORY OF THE CONTOUR LINK. NO ADVERSE EVENT ALLEGED. THE CUSTOMER WAS ASKED TO RETURN THE METER FOR INVESTIGATION. A REPLACEMENT METER WAS SENT TO HIM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 460998 | CONTOUR LINK | BLOOD GLUCOSE METER | NBW | BAYER HEALTHCARE LLC | 7778 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |