PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2015-74323
- Event Type
- Malfunction
- Date Received
- July 15, 2015
- Date of Event
- June 24, 2015
- Report Date
- June 24, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
THE INSULIN PUMP WAS RECEIVED WITH UNRESPONSIVE BUTTONS DUE TO UNLOCKED LCD CONNECTOR. THE INSULIN PUMP WAS ALSO RECEIVED WITH CRACKED RESERVOIR TUBE LIP.
IT WAS REPORTED THAT THE INSULIN PUMP HAD AN UNRESPONSIVE KEYPAD ON THE ESC BUTTON. THE CALLER DID NOT KNOW THE BLOOD GLUCOSE READING. THE CALLER WAS ADVISED TO REPLACE THE INSULIN PUMP AND A REQUEST WAS MADE TO HAVE THE DEVICE RETURNED FOR ANALYSIS.