PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2015-74199
- Event Type
- Malfunction
- Date Received
- July 14, 2015
- Date of Event
- June 24, 2015
- Report Date
- June 24, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL
- Reporter Occupation
- OTHER
Narratives
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE CUSTOMER REPORTED VIA PHONE CALL THAT SHE HAD A MOTOR ERROR ALARM YESTERDAY. DURING FILLING THE CANNULA, THE CUSTOMER TRIED AGAIN WITH A DIFFERENT SET AND RECEIVED A MOTOR POSITION ENCODER ERROR ALARM. CUSTOMER REWOUND AND SET THE TIME AND DATE ON THE PUMP. CUSTOMER RECEIVED A MOTION SENSOR TEST FAILURE ALARM. CUSTOMER'S BLOOD GLUCOSE WAS 300 MG/DL AT THE TIME OF THE INCIDENT. CUSTOMER STATED THAT THE PUMP WAS WORKING FINE UNTIL IT HAPPENED. CUSTOMER WAS ADVISED THAT THE PUMP WILL NEED TO BE REPLACED. CUSTOMER WAS ALSO ADVISED TO DISCONTINUE USE OF THE PUMP AND REVERT TO A BACK-UP PLAN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 456872 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR |