FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4914329 · Received July 14, 2015

Report

Report Number
3004209178-2015-74199
Event Type
Malfunction
Date Received
July 14, 2015
Date of Event
June 24, 2015
Report Date
June 24, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT SHE HAD A MOTOR ERROR ALARM YESTERDAY. DURING FILLING THE CANNULA, THE CUSTOMER TRIED AGAIN WITH A DIFFERENT SET AND RECEIVED A MOTOR POSITION ENCODER ERROR ALARM. CUSTOMER REWOUND AND SET THE TIME AND DATE ON THE PUMP. CUSTOMER RECEIVED A MOTION SENSOR TEST FAILURE ALARM. CUSTOMER'S BLOOD GLUCOSE WAS 300 MG/DL AT THE TIME OF THE INCIDENT. CUSTOMER STATED THAT THE PUMP WAS WORKING FINE UNTIL IT HAPPENED. CUSTOMER WAS ADVISED THAT THE PUMP WILL NEED TO BE REPLACED. CUSTOMER WAS ALSO ADVISED TO DISCONTINUE USE OF THE PUMP AND REVERT TO A BACK-UP PLAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
456872 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 59 YR