FDA Adverse Event Malfunction Summary report: N

ALLSET GOLD A HIGH RES SSP

MDR report key: 4912917 · Received July 10, 2015

Report

Report Number
2244574-2015-00095
Event Type
Malfunction
Date Received
July 10, 2015
Date of Event
May 22, 2015
Report Date
May 22, 2015
Manufacturer
LIFE TECHNOLOGIES
Product Code
MZI
PMA / PMN Number
BK070038
Removal / Correction Number
2244574-008-2015 C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION REPLICATED THE CUSTOMER'S OBSERVATION (B)(6) 2015. THE INVESTIGATION CONCLUDED ON (B)(6) 2015 AND WAS UNABLE TO DETERMINE WHETHER THE RESULT IS RELATED TO THE SPECIFIC DNA SAMPLE OR REFLECTIVE OF INCORRECT LABELING ON THE REACTIVITY FOR THE ALLELE. LIFE TECHNOLOGIES USED A RETAIN KIT AND TESTED A SAMPLE RETURNED BY THE CUSTOMER IN SSP ANALYSIS. RESULTS WERE OBTAINED ON (B)(6) 2015. THE CLINICAL CONSEQUENCES OF A MIS-TYPE OF THE AFFECTED ALLELES ARE MINIMAL TO NON-EXISTENT. IN ADDITION, THE AFFECTED ALLELE IS VERY RARE IN THE GENERAL POPULATION AND IS EXPECTED TO BE IN LESS THAN (B)(4) OF THE DNA SAMPLES TESTED.

Description of Event or Problem · 1

THIS REPORT IS RELATED TO CUSTOMER COMPLAINT (B)(4) MADE AGAINST ALLSET GOLD A HIGH RES SSP (SKU (B)(4), LOT 012 1608823), SINCE THIS LOT CONTAINS THE SAME AFFECTED PRIMER MIX. THE CUSTOMER REPORTED THAT THE REACTIVITY FOR LANE 86 WAS NEGATIVE FOR THE A03:20 ALLELE WHEN THE LABELING INDICATES THE REACTIVITY SHOULD BE POSITIVE. THE LABELING DISCREPANCY MAY LEAD TO POTENTIALLY MISTYPING AN A03:20 ALLELE AS AN A03:01 ALLELE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
448629 ALLSET GOLD A HIGH RES SSP MZI TEST, QUALITATIVE FOR HLA MZI LIFE TECHNOLOGIES 012 1615252

Patients

Seq Age Sex Outcome Treatment
1