FDA Adverse Event Malfunction Summary report: N

S3 UPS CHARGING MODULE

MDR report key: 4912808 · Received July 8, 2015

Report

Report Number
9611109-2015-00237
Event Type
Malfunction
Date Received
July 8, 2015
Date of Event
June 9, 2015
Report Date
February 22, 2017
Manufacturer
SORIN GROUP DEUTSCHLAND GMBH
Product Code
DTQ
PMA / PMN Number
K950990
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO PATIENT INVOLVEMENT. SORIN GROUP DEUTSCHLAND MANUFACTURES THE S3 UPS CHARGING MODULE. THE INCIDENT OCCURRED IN (B)(6). THIS MEDWATCH REPORT IS FILED ON BEHALF OF SORIN GROUP (B)(4). SORIN GROUP RECEIVED A REPORT THAT WHEN DISCONNECTED FROM THE MAINS, THE SYSTEM WOULD NOT SWITCH TO INTERNAL POWER UNLESS THE EMERGENCY DISCHARGE IS ENABLED. THERE WAS NO PATIENT INVOLVEMENT. THE INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SENT WHEN THE INVESTIGATION IS COMPLETE.

Additional Manufacturer Narrative · 1

LIVANOVA (B)(4) MANUFACTURES THE S3 UPS CHARGING MODULE. THE INCIDENT OCCURRED IN (B)(6). THIS MEDWATCH REPORT IS BEING FILED ON BEHALF OF LIVANOVA (B)(4). A LIVANOVA FIELD SERVICE REPRESENTATIVE WAS DISPATCHED TO THE FACILITY TO INVESTIGATE. THE SERVICE REPRESENTATIVE WAS ABLE TO CONFIRM THE REPORTED MALFUNCTION AND TRACED THE FAILURE TO A DEFECTIVE UPS CHARGE MODULE AND CONTROL MODULE. BOTH COMPONENTS WERE REPLACED TO RESOLVE THE ISSUE. SUBSEQUENT TESTING OF THE DEVICE FOUND NO FURTHER ISSUES. A REVIEW OF THE DHR DID NOT IDENTFY ANY DEVIATIONS OR NON-CONFORMITIES RELEVANT TO THE REPORTED ISSUE. LIVANOVA (B)(4) WILL CONTINUE TO MONITOR FOR TRENDS RELATED TO THIS TYPE OF ISSUE. EVALUATED ON SITE BY LIVANOVA TECHNICIAN.

Description of Event or Problem · 1

SORIN GROUP RECEIVED A REPORT THAT WHEN DISCONNECTED FROM THE MAINS, THE SYSTEM WOULD NOT SWITCH TO INTERNAL POWER UNLESS THE EMERGENCY DISCHARGE IS ENABLED. THERE WAS NO PATIENT INVOLVEMENT.