SAFETY SCALPEL
Report
- Report Number
- 3004605321-2015-00013
- Event Type
- Malfunction
- Date Received
- July 8, 2015
- Date of Event
- June 5, 2015
- Report Date
- July 7, 2015
- Manufacturer
- DEROYAL INTERCONTINENTAL, S.A.L
- Product Code
- GES
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
INVESTIGATION FINDINGS: THE QFI REPORT WAS REVIEWED FOR THE SALES AND SIMILAR COMPLAINT INFO. DEROYAL HAS SOLD (B)(4) CASES OF THE FINISHED GOOD FROM 2013 TO PRESENT. THERE HAVE BEEN NO PREVIOUS REPORTS IN WHICH AN ISSUE HAS BEEN PRESENT DUE TO NOT HAVING LABELING DETAILING THE MANNER IN WHICH THE BLADE EXTENDS. CORRECTION: A CORRECTION HAS NOT BEEN TAKEN. ROOT CAUSE ANALYSIS: THE END USER PERFORMED A KNOWLEDGE-BASED ERROR AND DID NOT FOLLOW THE INSTRUCTIONS FOR USE. CORRECTIVE ACTION AND/OR SYSTEMIC CORRECTION ACTION TAKEN: DUE TO THE INVESTIGATION AND ROOT CAUSE DETERMINATION, A CORRECTIVE ACTION HAS NOT BEEN TAKEN. PREVENTIVE ACTION: DUE TO THE INVESTIGATION AND ROOT CAUSE DETERMINATION, A PREVENTIVE ACTION HAS NOT BEEN TAKEN. THE INVESTIGATION IS COMPLETE AT THIS TIME. THIS REPORT WILL BE UPDATED IF MORE INFO BECOMES AVAILABLE.
CUSTOMER COMPLAINT RECEIVED FROM VNUS MEDICAL TECHNOLOGIES REGARDING THE SAFETY SCALPEL, 11 SS NS, PART #D4511A, SUPPLIER LOT # 38440124. A DOCTOR REPORTED CONCERNS WITH THE DESIGN OF THE SCALPEL. THIS ISSUE IS WHEN YOU SLIDE OPEN (IF YOU ARE NOT PAYING CLOSE ATTENTION), THERE IS NO LABEL STATING WHICH WAY TO OPEN THE SHARP END; AND IF IT'S TURNED OVER, YOU CANNOT TELL AS WELL. THIS IS NOT A LOT SPECIFIC COMPLAINT. THE DOCTOR CUT HIMSELF USING IT; HOWEVER, HE WAS ABLE TO CONTINUE DOING THE CASE BUT WAS CONCERNED FOR FUTURE PROCEDURES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 441599 | SAFETY SCALPEL | HANDLE, SCALPEL | GES | DEROYAL INTERCONTINENTAL, S.A.L | 38440124 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |