TERMINAL CUP IMPACTOR FOR METAL BACK CC
Report
- Report Number
- 3005180920-2015-00140
- Event Type
- Malfunction
- Date Received
- July 9, 2015
- Report Date
- October 7, 2015
- Manufacturer
- MEDACTA INTERNATIONAL, SA
- Product Code
- LXH
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
ON 07 SEPTEMBER 2015 IT WAS PREPARED A FINAL REPORT WITH THE INFORMATION ALREADY SUBMITTED IN THE INITIAL REPORT. ON THE SAME DAY THE REPORT WAS SENT TO THE INITIAL REPORTER AND THE CASE WAS CLOSED.
BATCH REVIEW PERFORMED ON (B)(6) 2015: LOT #1413789: (B)(4) INSTRUMENTS MANUFACTURED AND RELEASED ON (B)(4) 2014. NO ANOMALIES FOUND RELATED TO THE PROBLEM. LOT #150076: (B)(4) SHELLS MANUFACTURED AND RELEASED ON (B)(4) 2015. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) SHELLS OF THE LOT HAVE BEEN ALREADY SOLD. ON (B)(6) 2015, THE (B)(4) MADE THE FOLLOWING ANALYSIS UPON INSPECTION OF THE RETRIEVED ITEMS: THE TERMINAL WAS MANUALLY DISENGAGED FROM THE CUP WITH THE SCREWDRIVER USING A CLAMP TO HOLD THE CUP. THE CONTACT SURFACES BETWEEN THE TWO ITEMS DO NOT PRESENT SIGNS OF SEIZING OR EXCESSIVE WEAR.
(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 444712 | TERMINAL CUP IMPACTOR FOR METAL BACK CC | SURGICAL INSTRUMENT FOR HIP | LXH | MEDACTA INTERNATIONAL, SA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |