LOCATOR OVERDENTURE IMPLANT SYSTEM (LODI)
Report
- Report Number
- 2023950-2015-00095
- Event Type
- Injury
- Date Received
- July 9, 2015
- Report Date
- July 9, 2015
- Manufacturer
- ZEST ANCHORS, LLC
- Product Code
- DZE
- PMA / PMN Number
- K120198
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- NOT APPLICABLE
Narratives
USER DOCUMENTATION (TECHNIQUE MANUAL, P/N L8019-TM) SPECIFIES THAT IMMEDIATE LOADING IS SUITABLE ONLY IF SUFFICIENT PRIMARY STABILITY OF THE IMPLANT IS ACHIEVED AT THE TIME OF PLACEMENT. ALSO, THE RECOMMENDED IMPLANT INSERTION TORQUE AND IMPLANT ATTACHMENT TORQUE IS 30 N-CM. IF THE IMPLANT PLACEMENT TORQUE IS LESS THAN 30 N-CM, THEN THE ATTACHMENTS SHOULD ONLY BE HAND TIGHTENED. THE IMPLANT INSERTION TORQUE SHOULD NOT EXCEED 70 N-CM; IF 70 N-CM IS REACHED PRIOR TO FULL SEATING, THE IMPLANT SHOULD BE REMOVED AND THE OSTEOTOMY SHOULD BE ENLARGED. CLINICIAN STATED THAT LODI IMPLANTS FAILED TO OSSEOINTEGRATE. THE CLINICIAN DID NOT PROVIDE THE FOLLOWING INFORMATION: AMOUNT OF TORQUE USED ON ABUTMENT & IMPLANT, PATIENT BONE DENSITY, WHETHER PRIMARY STABILITY WAS ACHIEVED, WHETHER IMPLANT WAS IMMEDIATELY LOADED. FAILURE TO OSSEOINTEGRATE IS A WELL-DOCUMENTED INHERENT RISK OF DENTAL IMPLANTS. IN THE MAJORITY OF CASES WHERE AN IMPLANT FAILS TO INTEGRATE WITH THE BONE AND IS REJECTED BY THE BODY, THE CAUSE MAY BE UNK. THE IMPLANTS' LOT HISTORY RECORDS WERE REVIEWED; ANY DISCREPANCIES OR ISSUES OF NON-CONFORMANCE WERE SUCCESSFULLY RESOLVED PRIOR TO THE RELEASE OF THE PARTS. IMPLANT WAS MANUFACTURED TO SPECS. NO FURTHER ACTION IS REQUIRED.
LODI IMPLANTS FAILED TO OSSEOINTEGRATE. INCIDENT OCCURRED IN (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 447137 | LOCATOR OVERDENTURE IMPLANT SYSTEM (LODI) | ENDOSSEOUS DENTAL IMPLANT | DZE | ZEST ANCHORS, LLC | I0N2L |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |