FDA Adverse Event Other Summary report: N

BD LOGIC CO-BRANDED BLOOD GLUCOSE METER

MDR report key: 490780 · Received October 17, 2003

Report

Report Number
2243072-2003-00080
Event Type
Other
Date Received
October 17, 2003
Date of Event
September 17, 2003
Report Date
October 15, 2003
Manufacturer
BECTON DICKINSON
Product Code
CFR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

CONSUMER REPORTS HIGH READINGS WITH ONE VIAL OF STRIPS. ANALYSIS RUN ON CLARKE ERROR GRID USING COMPETITIVE METER READINGS (39,205) AS REF POINTS VS. BD READINGS (200, 425) YIELDED DATA POINTS IN THE E/C RANGES OF THE GRID, OUTSIDE ACCEPTABLE RANGES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD LOGIC CO-BRANDED BLOOD GLUCOSE METER BD TEST STRIPS CFR BECTON DICKINSON NA 2071325

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other