FREESTYLE LITE
Report
- Report Number
- 2954323-2015-00348
- Event Type
- Malfunction
- Date Received
- July 10, 2015
- Date of Event
- June 3, 2015
- Report Date
- October 19, 2015
- Manufacturer
- ABBOTT DIABETES CARE INC.
- Product Code
- NBW
- PMA / PMN Number
- K092602
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
THE REPORTED METER WAS RETURNED AND INVESTIGATED WITH RETAINED TEST STRIPS. THE COMPLAINT IS NOT CONFIRMED AND NO NEW ISSUES WERE OBSERVED. ALL RESULTS WERE WITHIN THE RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED DURING CONTROL SOLUTION TESTING. THE REPORTED READINGS WERE FOUND IN THE METER'S MEMORY.
THE PRODUCT HAS BEEN REQUESTED BACK FOR INVESTIGATION. A FOLLOW- UP REPORT WILL BE SUBMITTED ONCE ADDITIONAL INFORMATION IS OBTAINED. ALL PERTINENT INFORMATION AVAILABLE TO ABBOTT DIABETES CARE HAS BEEN SUBMITTED.
GIVEN NO TEST STRIPS WERE RETURNED FOR INVESTIGATION, RETAINED TEST STRIP SAMPLES FROM THE SAME LOT REPORTED BY THE CUSTOMER ((B)(4)) WERE TESTED WITH CONTROL SOLUTION. ALL RESULTS WERE WITHIN RANGE SPECIFICATION AND NO ERRORS WERE OBSERVED. THE COMPLAINT IS NOT CONFIRMED.
CUSTOMER REPORTED RECEIVING ERRATIC READINGS ON THEIR ADC BLOOD GLUCOSE METER. CUSTOMER REPORTED RECEIVING READINGS OF 2.0 MMOL/L AND 19.7 MMOL/L WITHIN 10 MINUTES. THE RESULTS WHEN PLOTTED ON A PARKES ERROR GRID FELL INTO THE "C" ZONE SHOWING THE DIFFERENCE IN VALUES TO BE CLINICALLY SIGNIFICANT. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY, OR MISTREATMENT ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 449101 | FREESTYLE LITE | BLOOD GLUCOSE MONITORING SYSTEM | NBW | ABBOTT DIABETES CARE INC. | 1474612 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |