FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 4905686
·
Received July 7, 2015
Report
- Report Number
- 1314492-2015-07321
- Event Type
- Malfunction
- Date Received
- July 7, 2015
- Date of Event
- June 1, 2015
- Report Date
- June 13, 2015
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
(B)(4). BAXTER RECEIVED AND EVALUATED THE DEVICE. THE DEVICE WAS FOUND OUT OF SPECIFICATION IN RELATION TO THE SYSTEM ERROR 105, WHICH WAS REPRODUCED DURING POWER UP. SYSTEM ERROR 105 WAS CONFIRMED AND CAUSED BY A FAILED MOTOR FLEX. THE FAILED MOTOR ASSEMBLY WAS REPLACED. SEE SCANNED PAGE.
Description of Event or Problem · 1
IT WAS REPORTED THAT A SPECTRUM PUMP DISPLAYED SYSTEM ERROR 105. IT WAS ALSO REPORTED THERE WAS NO PT INVOLVEMENT.