POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI
Report
- Report Number
- 3008262382-2015-01477
- Event Type
- Malfunction
- Date Received
- July 10, 2015
- Date of Event
- June 16, 2015
- Report Date
- July 29, 2015
- Manufacturer
- INVACARE REHABILITATION EQUIPMENT CO.
- Product Code
- ITI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL RECORD WILL BE FILED.
(B)(4). EXPANDED EVALUATION REPORT RECEIVED 7.29.2015. THE CAUSE OF MECHANICALLY WEAK BRAKE COULD NOT BE DETERMINED. ONE POSSIBLE CAUSE COULD BE A LAYER OF THE BRAKE COIL BEING SLIGHTLY OUT OF SPECIFICATION, CAUSING AN ERROR IN THE SPACING OF COMPONENTS. CONCLUSIONS: UTILIZING EXISTING COMPLAINT INFORMATION, ACTUAL OBSERVATIONS, AND FUNCTIONAL TESTING OF THE RETURNED PRODUCT IN ITS "AS RECEIVED" CONDITION, THE COMPLAINT WAS CONFIRMED FOR THE 1134124 RIGHT MOTOR ASSEMBLY BRAKE TO NOT ENGAGE FULLY. THE ELECTRICAL SYSTEM ENGAGES THE BRAKE PROPERLY, BUT WHEN ENGAGED THE BRAKE WAS NOT STRONG ENOUGH TO ADEQUATELY HOLD THE MOTOR SHAFT STATIONARY.
PROVIDER STATES THE RIGHT MOTOR BRAKE WILL NOT ENGAGE, OUT OF BOX.
PROVIDER STATES THE RIGHT MOTOR BRAKE WILL NOT ENGAGE, OUT OF BOX.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 449004 | POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI | WHEELCHAIR, POWERED | ITI | INVACARE REHABILITATION EQUIPMENT CO. | M41SRR |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |