FDA Adverse Event Malfunction Summary report: N

POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI

MDR report key: 4905227 · Received July 10, 2015

Report

Report Number
3008262382-2015-01477
Event Type
Malfunction
Date Received
July 10, 2015
Date of Event
June 16, 2015
Report Date
July 29, 2015
Manufacturer
INVACARE REHABILITATION EQUIPMENT CO.
Product Code
ITI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL RECORD WILL BE FILED.

Additional Manufacturer Narrative · 1

(B)(4). EXPANDED EVALUATION REPORT RECEIVED 7.29.2015. THE CAUSE OF MECHANICALLY WEAK BRAKE COULD NOT BE DETERMINED. ONE POSSIBLE CAUSE COULD BE A LAYER OF THE BRAKE COIL BEING SLIGHTLY OUT OF SPECIFICATION, CAUSING AN ERROR IN THE SPACING OF COMPONENTS. CONCLUSIONS: UTILIZING EXISTING COMPLAINT INFORMATION, ACTUAL OBSERVATIONS, AND FUNCTIONAL TESTING OF THE RETURNED PRODUCT IN ITS "AS RECEIVED" CONDITION, THE COMPLAINT WAS CONFIRMED FOR THE 1134124 RIGHT MOTOR ASSEMBLY BRAKE TO NOT ENGAGE FULLY. THE ELECTRICAL SYSTEM ENGAGES THE BRAKE PROPERLY, BUT WHEN ENGAGED THE BRAKE WAS NOT STRONG ENOUGH TO ADEQUATELY HOLD THE MOTOR SHAFT STATIONARY.

Description of Event or Problem · 1

PROVIDER STATES THE RIGHT MOTOR BRAKE WILL NOT ENGAGE, OUT OF BOX.

Description of Event or Problem · 1

PROVIDER STATES THE RIGHT MOTOR BRAKE WILL NOT ENGAGE, OUT OF BOX.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
449004 POWER CHAIR 20 WIDE BY 18 DEEP FOR PMI WHEELCHAIR, POWERED ITI INVACARE REHABILITATION EQUIPMENT CO. M41SRR

Patients

Seq Age Sex Outcome Treatment
1 Other