PUREPOINT LASER
Report
- Report Number
- 2028159-2015-06923
- Event Type
- Malfunction
- Date Received
- July 10, 2015
- Date of Event
- March 18, 2015
- Report Date
- August 13, 2015
- Manufacturer
- ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
- Product Code
- HQF
- PMA / PMN Number
- K062624
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITION REPORTABLE INFORMATION BECOMES AVAILABLE. (B)(4).
THE SYSTEM WAS EXAMINED AND THE REPORTED EVENT WAS NOT REPLICATED. THE SYSTEM WAS TESTED AND FOUND TO MEET PRODUCT SPECIFICATIONS. A REVIEW OF THE CUSTOMER¿S COMPLAINT HISTORY FOR THE LAST 24 MONTHS DID NOT SHOW ANY PREVIOUS COMPLAINTS OF THIS KIND AGAINST THE SYSTEM. THE SYSTEM WAS MANUFACTURED ON APRIL 26, 2013. BASED ON QA ASSESSMENT, THE PRODUCT MET SPECIFICATIONS AT THE TIME OF RELEASE. THE SYSTEM WAS FOUND TO MEET SPECIFICATIONS. THEREFORE, THE ROOT CAUSE OF THE REPORTED EVENT CANNOT BE DETERMINED CONCLUSIVELY. (B)(4).
A FILED SERVICE TECHNICIAN NOTICED DURING ROUTINE MAINTENANCE THAT THE LASER OUTPUT POWER WAS A LITTLE HIGH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 449554 | PUREPOINT LASER | LASER, OPHTHALMIC | HQF | ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |