FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4904193 · Received July 9, 2015

Report

Report Number
3004209178-2015-72681
Event Type
Malfunction
Date Received
July 9, 2015
Date of Event
June 19, 2015
Report Date
June 19, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD NO BUTTON RESPONSE DUE TO MOISTURE DAMAGE ON KEYPAD TRACES. OCCLUSION TEST WAS NOT PERFORMED DUE TO BUTTON/KEYPAD ANOMALY. NO MOISTURE DAMAGE WAS FOUND INSIDE OF THE DEVICE. THE DEVICE DID HAVE CRACKED CASE AT DISPLAY WINDOW CORNER, RESERVOIR TUBE LIP, AND MINOR SCRATCHED DISPLAY WINDOW.

Description of Event or Problem · 1

CUSTOMER REPORTED VIA PHONE, SHE BELIEVES THERE IS AN ISSUE WITH THE KEYPAD, SINCE THE BUTTONS DO NOT RESPOND WHEN SHE PRESSES THEM. CUSTOMER'S BLOOD GLUCOSE WAS 275 MG/DL. ISSUE IS OCCURRING WITH ALL OF THE BUTTONS AND WHEN SHE PRESSES THE B BUTTON IT DISPLAYS LOW RESERVOIR. THE DEVICE WAS ACCIDENTLY WET, FELL INTO THE SINK, AND IT MAY HAVE CAUSED THE KEYPAD ISSUES. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE, REVERT TO BACK UP PLAN, AND THE DEVICE NEED TO BE REPLACED. SHE AGREED TO RETURN THE DEVICE FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
444897 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523LNAL

Patients

Seq Age Sex Outcome Treatment
1 60 YR