FDA Adverse Event Malfunction Summary report: N

16" EAGLE 2000 GRAVITY STERILIZER

MDR report key: 490412 · Received October 10, 2003

Report

Report Number
2515984-2003-00001
Event Type
Malfunction
Date Received
October 10, 2003
Date of Event
September 10, 2003
Report Date
September 10, 2003
Manufacturer
STERIS CORPORATION - ERIE
Product Code
FLE
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

UNIT WAS ON BUT NOT HEATING. MAINTENANCE FOUND THE THREE PHASE BREAKER WAS TRIPPED. WHEN THE MAINTENANCE RESET IT THE OPERATOR WENT TO SET THE SWITCH FROM 'STANDBY TO 'ON'. AT THIS TIME THE OPERATOR FELT A SHOCK AND WITNESSED A BRIGHT NOISY FLASH. THE OPERATOR GOT A TINGLE IN THEIR FINGERS WHICH CAUSED THEM TO TOUCH THE COUNTERTOP BEHIND. STERIS TECHNICIAN FOUND THE HEATER COIL WAS BURNT OUT AND THE GROUND WIRE FROM THE CHASSIS TO HEATER PLATE WAS BURNT AND CONTRACTOR POINTS BLACKENED. TECH REPLACED HEATER ASSY AND CONTACTOR AND REPAIRED GROUND WIRE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 16" EAGLE 2000 GRAVITY STERILIZER STEAM STERILIZER FLE STERIS CORPORATION - ERIE 2011 NA

Patients

Seq Age Sex Outcome Treatment
1 * Other