FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4904111 · Received July 9, 2015

Report

Report Number
3004209178-2015-73021
Event Type
Malfunction
Date Received
July 9, 2015
Date of Event
June 20, 2015
Report Date
June 20, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED FOR A BUTTON ERROR AND HAD INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BLANK DISPLAY WAS NOTED. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON THE DISPLAY WINDOW AND A SCRATCHED RESERVOIR TUBE WINDOW.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP ALARMED WITH A BUTTON ERROR, AFTER SHE TESTED HER BLOOD GLUCOSE AND THE SCREEN WENT BLANK FOR TEN MINUTES. CUSTOMER'S BLOOD GLUCOSE WAS 247 MG/DL. THE INSULIN PUMP HAD BEEN EXPOSED TO HUMIDITY. IT WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO A BACK-UP PLAN. IT WAS FURTHER ADVISED THAT THE DEVICE WOULD BE REPLACED AND AGREED UPON THAT THE PRODUCT WOULD BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
445438 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAL

Patients

Seq Age Sex Outcome Treatment
1 58 YR