FDA Adverse Event Malfunction Summary report: N

ADVANTX LEGACY CLASSICAL RADIOGRAPHY AND FLUOROSCOPY S

MDR report key: 490393 · Received October 22, 2003

Report

Report Number
490393
Event Type
Malfunction
Date Received
October 22, 2003
Date of Event
August 1, 2003
Report Date
August 1, 2003
Manufacturer
GENERAL ELECTRIC MEDICAL SYSTEMS
Product Code
JAA
Product Problem
Yes
Report Source
User Facility report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

A PT WAS ON A FLUORO TABLE FOR ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY. THE FLUORO MACHINE/TUBE WARMED UP AS PER PROCEDURE. AFTER THE PT WAS SEDATED, FLUORO WAS PULLED INTO PLACE AND MALFUNCTIONED. IMAGES WERE UNABLE TO BE SEEN. BIOMEDICAL SERVICES WERE CALLED TO REPAIR THE MACHINE. THE PT WAS MOVED TO RECOVERY AND THE PROCEDURE WAS DONE LATER THAT AFTERNOON. GE MEDICAL SYSTEMS FOUND A BAD POWER SUPPLY, WHICH WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTX LEGACY CLASSICAL RADIOGRAPHY AND FLUOROSCOPY S RADIOGRAPHY AND FLUOROSCOPY JAA GENERAL ELECTRIC MEDICAL SYSTEMS * *

Patients

Seq Age Sex Outcome Treatment
1 40 YR Other