PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2015-03839
- Event Type
- Injury
- Date Received
- July 9, 2015
- Date of Event
- May 29, 2015
- Report Date
- February 29, 2016
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). IT IS INDICATED THAT THE DEVICE IS NOT RETURNING FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE INVESTIGATION WAS UNABLE TO DETERMINE A CONCLUSIVE CAUSE FOR THE REPORTED DIFFICULTIES; HOWEVER, THE TREATMENT APPEARS TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A LOT SPECIFIC QUALITY ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING.
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED NEEDLE TO CUFF MISS/SUTURE RETRIEVAL ISSUE WAS CONFIRMED. IN ADDITION, ANALYSIS FOUND ONE ANTERIOR CUFF TAB WAS DETACHED AND NOT RETURNED WITH THE DEVICE. CUFF TABS MEASURE ONE ONE-HUNDREDTH (0.01) OF AN INCH SQUARE. DUE TO THE EXTREMELY SMALL SIZE, A MISSING CUFF TAB IS NOT VISIBLE WITHOUT MAGNIFICATION AND WOULD NOT BE NOTED BY THE USER. BASED ON VISUAL AND FUNCTIONAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING. THE INVESTIGATION DETERMINED THE REPORTED DIFFICULTIES AND TREATMENT APPEAR TO BE RELATED TO CIRCUMSTANCES OF THE PROCEDURE. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO MANUFACTURING NONCONFORMITIES THAT WOULD HAVE CONTRIBUTED TO THIS COMPLAINT. ADDITIONALLY, A REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A LOT SPECIFIC QUALITY ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO MANUFACTURE, DESIGN OR LABELING.
(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION. THE OTHER FOUR PROGLIDE DEVICES REFERENCED ARE FILED UNDER SEPARATE MEDWATCH MFR REFERENCE NUMBERS.
SUBSEQUENT TO THE INITIAL MEDWATCH REPORT, THE FOLLOWING ADDITIONAL INFORMATION WAS RECEIVED: IT WAS REPORTED THAT SUTURE PLACEMENT IN THE RIGHT COMMON FEMORAL ARTERY WAS ATTEMPTED WITH PROGLIDE DEVICES, USING THE PRE-CLOSE TECHNIQUE, VIA AN 8F SHEATH PRIOR TO AN ABDOMINAL AORTIC ANEURYSM (AAA) PROCEDURE. NO FEMORAL ANGIOGRAM WAS PERFORMED. REPORTEDLY, AFTER PLUNGER REMOVAL, THE SUTURES WERE NOT CONNECTED TO THE DEVICE AND THE KNOT COULD NOT BE FORMED. FOUR ADDITIONAL PROGLIDE DEVICES WERE USED WITH THE SAME RESULTS. THE MAXIMUM SHEATH SIZE WAS 8F. THE AAA PROCEDURE WAS COMPLETED AND HEMOSTASIS WAS ACHIEVED BY APPLYING MANUAL ARTERIAL COMPRESSION. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. IT WAS REPORTED THAT THE PHYSICIAN IS TRAINED IN THE USE OF THE PROGLIDE DEVICE AND ESTABLISHED IN THE PRECLOSE TECHNIQUE. NO ADDITIONAL INFORMATION WAS PROVIDED.
IT WAS REPORTED THAT CLOSURE OF AN UNKNOWN VESSEL WAS ATTEMPTED USING A PROGLIDE DEVICE AFTER AN ABDOMINAL AORTIC ANEURYSM (AAA) PROCEDURE. REPORTEDLY, AFTER PLUNGER REMOVAL, THE SUTURES WERE NOT CONNECTED TO THE DEVICE AND THE KNOT COULD NOT BE FORMED. FOUR ADDITIONAL PROGLIDE DEVICES WERE USED WITH THE SAME RESULTS. MANUAL ARTERIAL COMPRESSION WAS APPLIED TO ACHIEVE HEMOSTASIS. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. IT IS UNKNOWN IF THE PHYSICIAN IS TRAINED IN THE USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 445162 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | AV-TEMECULA-CT | 40331K1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Required Intervention |