VERTEX RECONSTRUCTION SYSTEM
Report
- Report Number
- 1030489-2015-01425
- Event Type
- Injury
- Date Received
- July 9, 2015
- Date of Event
- June 10, 2015
- Report Date
- November 20, 2015
- Manufacturer
- WARSAW ORTHOPEDICS
- Product Code
- KWP
- PMA / PMN Number
- K121191
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). NEITHER DEVICE NOR APPLICABLE IMAGING RECEIVED. THEREFORE CAUSE OF EVENT CANNOT BE DETERMINED.
PRODUCT ANALYSIS: VISUAL AND OPTICAL EXAMINATION OF THE SET SCREW APPEAR TO DISPLAY ASYMMETRICAL WITNESS MARKS ON ONE OF THE COMPLAINT RETURN SET SCREWS. THIS IS CONSISTENT WITH INCOMPLETE SEATING OF THE SET SCREW TO THE ASSOCIATED ROD.
IT WAS REPORTED THAT THE PATIENT UNDERWENT SURGERY AND THE SURGEON ADDED SCREWS AT T1 <(>&<)> T2 TO THE EXISTING C3-C6 CONSTRUCT. THE SURGEON THEN WENT ON TO REMOVE THE C6 SCREW ON THE LEFT SIDE TO ACCOMMODATE AN AXIAL DOMINO CONNECTOR. AT DAY 3,POST-OP, X-RAYS REVEALED THE AXIAL DOMINO HAD "POPPED OFF" ACCORDING TO THE SURGEON WHO CALLED THE PATIENT BACK TWO DAYS LATER TO REMOVE THE AXIAL DOMINO AND REVISE. ON THE RIGHT THE SURGEON USED A SIDE TO SIDE DOMINO CONNECTOR, WHICH REMAINS INTACT. THE INSTRUMENT DID NOT BREAK. THERE WERE PATIENT COMPLICATIONS AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 448290 | VERTEX RECONSTRUCTION SYSTEM | APPLIANCE, FIXATION, SPINAL INTERLAMINAL | KWP | WARSAW ORTHOPEDICS | NA | 0355824W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |