FDA Adverse Event Malfunction Summary report: N

TOTALCARE BARIATRIC BED

MDR report key: 4902694 · Received July 1, 2015

Report

Report Number
1824206-2015-00698
Event Type
Malfunction
Date Received
July 1, 2015
Date of Event
June 8, 2015
Report Date
June 8, 2015
Product Code
FNL
PMA / PMN Number
K122473
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACCOUNT CALLED TO GET THE PART NUMBER FOR THE SIDE COMMUNICATION CABLE. SEE SCANNED PAGE. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNK IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE ACCOUNT REPLACED THE COMMUNICATION CABLE TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM REC'D A REPORT FROM THE ACCOUNT STATING THE NURSE CALL WOULD NOT WORK. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
429541 TOTALCARE BARIATRIC BED TOTALCARE BARIATRIC BED FNL 1840

Patients

Seq Age Sex Outcome Treatment
1