PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4901848 · Received July 9, 2015

Report

Report Number
3004209178-2015-72617
Event Type
Malfunction
Date Received
July 9, 2015
Date of Event
June 7, 2015
Report Date
June 18, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD A BLANK DISPLAY DUE TO A BATTERY TUBE CONNECTOR NOT PROPERLY PLUGGED IN TO THE INTERFACE BOARD. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNER, MINOR SCRATCHES ON THE DISPLAY WINDOW, A SCRATCHED RESERVOIR TUBE WINDOW, CRACKED BELT CLIP SLOT, CRACKED BATTERY TUBE THREADS AND A CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER HAD A POWER SWITCH DEFECT ON THE INSULIN PUMP. THE PUMP WILL NOT WORK EVEN WITH A NEW BATTERY. THE INSULIN PUMP WILL BE REPAIRED AND REPLACED.