ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2015-22845
- Event Type
- Malfunction
- Date Received
- July 8, 2015
- Report Date
- June 22, 2015
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVALUATION. AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1: DATE OF SUBMISSION 9/3/15 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2015 WITH THE FOLLOWING FINDINGS: PUMP WAS RETURNED WITH MOISTURE IN THE BATTERY COMPARTMENT. LEAK TEST FAILED DUE TO A CRACK ALONGSIDE THE BATTERY COMPARTMENT. A BLANK DISPLAY SCREEN WAS OBSERVED. OPENED PUMP- MOISTURE FOUND ON OLED, PCB, DISPLAY FLEX AND FLEX CONNECTOR. UNABLE TO VERIFY ALL STEPS DUE TO BLANK SCREEN.
ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A CASING/CONDITION (MOISTURE INGRESS) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 443043 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |