FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4900799 · Received July 8, 2015

Report

Report Number
2531779-2015-22818
Event Type
Malfunction
Date Received
July 8, 2015
Report Date
June 21, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVALUATION. AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 8/18/2015 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2015 WITH THE FOLLOWING FINDINGS: A "BATTERY LIFE" ISSUE WAS NOT DUPLICATED DURING INVESTIGATION. BLACK BOX SHOWS EVIDENCE OF BATTERY ALARMS AND SHORT BATTERY LIFE. CURRENT DRAWS WITHIN SPECIFICATIONS. BATTERY CAP IS ABLE TO FULLY TIGHTEN HOWEVER BATTERY CAP THREADS DAMAGED. BATTERY COMPARTMENT CRACK OBSERVED BELOW GRIP PAD. EVIDENCE OF MOISTURE CONTAMINATION INSIDE BATTERY COMPARTMENT AND UNDERSIDE OF BATTERY CAP. UNRELATED TO THE COMPLAINT, THE PUMP POWERS WITH RETURNED BATTERY CAP TO A DIM DISCOLORED DISPLAY. LEAK TEST PASSES. REMOVED PUMP CASE AND DISASSEMBLED PUMP. EVIDENCE OF ADDITIONAL MOISTURE CONTAMINATION INSIDE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
442934 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 55 YR