ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2015-22804
- Event Type
- Malfunction
- Date Received
- July 8, 2015
- Report Date
- June 21, 2015
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- 003
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1: DATE OF SUBMISSION 29-MAY-2019. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 23-MAY-2019 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THERE WAS MOISTURE OBSERVED BEHIND THE DISPLAY LENS. A LEAK TEST FOUND A LEAK FROM A CRACK BETWEEN DISPLAY LENS AND CASE SEAL. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THIS REPORT IS MADE UNDER THE REQUIREMENTS OF THE MEDICAL DEVICE REPORTING REGULATIONS AND DOES NOT CONSTITUTE AN ADMISSION ON THE PART OF ANIMAS OF ANY DEFICIENCY IN THE PERFORMANCE OF THE DEVICE.
ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (CLOUDY W/ MOISTURE) ISSUE; MOISTURE BEHIND THE DISPLAY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 442935 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |