FDA Adverse Event Malfunction Summary report: N

LOCKING SCREW VARIAX FULL THREAD 3.5MM / L18MM

MDR report key: 4900186 · Received July 8, 2015

Report

Report Number
0008031020-2015-00283
Event Type
Malfunction
Date Received
July 8, 2015
Date of Event
June 11, 2015
Report Date
March 1, 2017
Manufacturer
STRYKER LEIBINGER FREIBURG
Product Code
HRS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ONCE THE INVESTIGATION HAS BEEN COMPLETED, ANY ADDITIONAL INFORMATION WILL BE REPORTED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

THE ITEM WAS MISIDENTIFIED AT THE TIME THE COMPLAINT WAS OPENED. THE REPORTED EVENT OF LOCKING SCREW VARIAX FULL THREAD 3.5MM / L18MM , DEVICE NOT ENGRAVED, COULD NOT BE CONFIRMED. VISUAL INSPECTION ON THE RETURNED DEVICES SHOWS THAT THE DEVICE WAS RECEIVED COMPLETE AND INTACT. DIMENSIONAL INSPECTION ON THE RETURNED DEVICES SHOWS THE DEVICE MANUFACTURED PER SPECIFICATIONS. NO DEVIATION FROM THE DRAWING KE 1904500 WAS FOUND. THE DEVICE IS NOT SUPPOSED TO BE ENGRAVED. NO DEVICE FAILURE/ANOMALY WAS IDENTIFIED. A REVIEW OF THE DEVICE HISTORY WAS NOT POSSIBLE BECAUSE THE LOT NUMBER WAS NOT COMMUNICATED. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. A REVIEW OF THE LABELING DID NOT INDICATE ANY ABNORMALITIES. NO CORRECTIVE ACTIONS ARE REQUIRED AT THIS TIME. NO INDICATIONS OF MATERIAL, MANUFACTURING OR DESIGN RELATED PROBLEMS WERE FOUND DURING THE INVESTIGATION.

Description of Event or Problem · 1

DURING CONTROL AND INSPECTION OF THE VARIAX KITS WITH THE CUSTOMER, SALES REP, OBSERVED THAT THERE WAS NO LOT NUMBER ON SOME OF THE LOCKING SCREWS.

Description of Event or Problem · 1

DURING CONTROL AND INSPECTION OF THE VARIAX KITS WITH THE CUSTOMER, SALES REP, OBSERVED THAT THERE WAS NO LOT NUMBER ON SOME OF THE LOCKING SCREWS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
444162 LOCKING SCREW VARIAX FULL THREAD 3.5MM / L18MM PLATE, FIXATION, BONE HRS STRYKER LEIBINGER FREIBURG UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Other