LTV
Report
- Report Number
- 2031702-2015-00126
- Event Type
- Malfunction
- Date Received
- July 8, 2015
- Report Date
- July 8, 2015
- Manufacturer
- CAREFUSION 203, INC
- Product Code
- CBK
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
THE MANUFACTURER WAS UNABLE TO DUPLICATE THE REPORTED PROBLEM. THE VENTILATOR PASSED AN EXTENDED TEST RUN USING THE CUSTOMER¿S VENTILATOR SETTINGS. THE VENTILATOR PASSED THE LTV FINAL TEST WHICH INCLUDES MANY VENTILATION AND ALARM FUNCTIONS. THE VENTILATOR IS NOT DESIGNED TO ALARM FOR A DISC/SENSE CONDITION WHEN THE PATIENT CIRCUIT IS DISCONNECTED DOWNSTREAM FROM THE PATIENT CIRCUIT WYE. PER THE OPERATOR'S MANUAL, THE DISC/SENSE ALARM DETECTS WHEN THE PATIENT CIRCUIT WYE ¿SENSE LINES¿ ARE NOT CONNECTED TO THE VENTILATOR. IT IS NOT DESIGNED TO DETECT A ¿PATIENT DISCONNECT¿ CONDITION. AS LONG AS DOWNSTREAM FLOW IS DETECTED THROUGH THE PATIENT CIRCUIT WYE, NO DISC/SENSE ALARM CONDITION WILL BE ISSUED. THE ¿LOW PRESSURE¿ AND ¿LOW MINUTE VOLUME¿ ALARM SETTINGS, WHEN SET APPROPRIATELY, ARE PROVIDED TO DETECT THE ABSENCE OF PROPER PATIENT VENTILATION. THE EVENT TRACE DID NOT CONTAIN ANY UNUSUAL ALARM TYPES OR INCIDENT COUNTS. ALL EVENT TRACE ENTRIES ARE CONSISTENT WITH NORMAL VENTILATOR OPERATION.
IT WAS REPORTED THAT THE VENTILATOR DID NOT ALARM WHEN THE CIRCUIT WAS DISCONNECTED. THE REPORTED EVENT OCCURRED TWICE. IT IS UNKNOWN IF THE PATIENT CIRCUIT WAS DISCONNECTED AT THE PATIENT WYE OR AT THE VENTILATOR. NO PATIENT HARM REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 443870 | LTV | VENTILATOR, CONTINUOUS | CBK | CAREFUSION 203, INC | LTV 1150 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |