FDA Adverse Event Malfunction Summary report: N

HOMEPUMP C-SERIES

MDR report key: 4898989 · Received July 1, 2015

Report

Report Number
MW5043697
Event Type
Malfunction
Date Received
July 1, 2015
Date of Event
June 29, 2015
Report Date
July 1, 2015
Manufacturer
KIMBERLY CLARK
Product Code
MEB
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
NJ, US
Reporter Occupation
PHARMACIST

Narratives

Description of Event or Problem · 1

WE SET UP A 5-FU INFUSION USING THE HOMEPUMP C- SERIES ELASTOMERIC PUMP PRODUCED BY I-FLOW. THE INFUSION BEGAN ON (B)(6) 2015 AT 2100 HOURS AND RAN DRY AT (B)(6) 2015 AT 0400. THE TOTAL INFUSION TIME IS SUPPOSED TO BE 46 HOURS AND THE INFUSION RAN ABOUT 31-32 HOURS. THIS IS THE 3RD INSTANCE OF THIS OCCURRING. WE PERFORMED OUR OWN INTERNAL AUDIT UTILIZING JUST SALINE AND THE TOTAL INFUSION TIME WAS ABOUT 41 HOURS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
426598 HOMEPUMP C-SERIES HOMEPUMP C-SERIES MEB KIMBERLY CLARK C-SERIES 0201617874

Patients

Seq Age Sex Outcome Treatment
1 75 YR