ALTRUA
Report
- Report Number
- 2124215-2015-08653
- Event Type
- Injury
- Date Received
- July 7, 2015
- Date of Event
- June 1, 2015
- Report Date
- July 14, 2015
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWP
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, DETAILED MECHANICAL AND ELECTRICAL TESTING WAS PERFORMED ON THE DEVICE. THE DEVICE BATTERY STATUS WAS EOL. THE DEVICE FUNCTIONED NORMALLY THROUGHOUT TESTING. LABORATORY ANALYSIS DETERMINED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION; HOWEVER, THE ESTIMATED LONGEVITY REMAINING VALUE APPEARED TO DECREASE MORE QUICKLY THAN EXPECTED BETWEEN ROUTINE FOLLOW-UPS. FACTORS INFLUENCING THE ESTIMATED LONGEVITY REMAINING CALCULATION INCLUDE PACING RATE, AMPLITUDE, PULSE-WIDTH AND LEAD IMPEDANCE. ANY (EVEN SLIGHT) CHANGES IN THESE FACTORS WILL IMPACT THE BATTERY CONSUMPTION CALCULATION AND THEREFORE THE REMAINING LONGEVITY ESTIMATE. PLEASE NOTE THAT, DESPITE THE DROP IN ESTIMATED LONGEVITY REMAINING, THE ACTUAL BATTERY CONDITION DID NOT CHANGE SIGNIFICANTLY BETWEEN FOLLOW-UPS. IN SUMMARY, IT WAS DETERMINED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION, BUT DECLARED EOL EARLIER THAN PREVIOUSLY ESTIMATED.
(B)(4). THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER DECLARED END OF LIFE (EOL) FASTER THAN EXPECTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THE DEVICE WAS EXPLANTED WITHOUT INCIDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 440866 | ALTRUA | IMPLANTABLE PULSE GENERATOR | LWP | GUIDANT CRM CLONMEL IRELAND | S502 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 81 YR | Hospitalization| L| R |