TRUERESULT
Report
- Report Number
- 1052693-2015-01048
- Event Type
- Malfunction
- Date Received
- July 1, 2015
- Date of Event
- June 8, 2015
- Report Date
- December 15, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
(B)(4). MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION USER HAD AN INACCURATE REFERENCE. INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND.
(B)(4). PRODUCT UNDER EVALUATION.
CONSUMER COMPLAINT ABOUT HIGH BLOOD RESULTS. CUSTOMER STATES HE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER'S EXPECTED BLOOD RESULTS ARE 90-130MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 04/23/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WERE FIRST OPENED (B)(6) 2015. CUSTOMER PERFORMED BACK TO BACK BLOOD TEST, 94MG/DL AND 94MG/DL FASTING. REVIEWED METER MEMORY: 1: 192MG/DL (B)(6) 2015 09:28:00 PM FASTING: YES. 2: 119MG/DL (B)(6) 2015 09:28:00 PM FASTING: YES. 3: 119MG/DL (B)(6) 2015 09:01:00 PM FASTING: YES. 4: 103MG/DL (B)(6) 2015 08:41:00 PM FASTING: YES. 5: 190MG/DL (B)(6) 2015 07:05:00 PM FASTING: YES. ADVERSE EVENT NOT REPORTED.
CONSUMER COMPLAINT ABOUT HIGH BLOOD RESULTS. CUSTOMER STATES THAT HE FEELS WELL AND REQUIRES NO MEDICAL ATTENTION. CUSTOMER'S EXPECTED BLOOD RESULTS ARE 90-130MG/DL FASTING. VERIFIED THE STRIPS EXPIRE 04/23/2018. CUSTOMER CONFIRMS THE STRIPS ARE STORED PROPERLY AND WERE FIRST OPENED (B)(6) 2015. CUSTOMER PERFORMED BACK TO BACK BLOOD TEST, 94MG/DL AND 94MG/DL FASTING. REVIEWED METER MEMORY: 192MG/DL, (B)(6) 2015 09:28:00 PM, FASTING: YES. 119MG/DL, (B)(6) 2015 09:28:00 PM, FASTING: YES. 119MG/DL, (B)(6) 2015 09:01:00 PM, FASTING: YES. 103MG/DL, (B)(6) 2015 08:41:00 PM, FASTING: YES. 190MG/DL, (B)(6) 2015 07:05:00 PM, FASTING: YES. ADVERSE EVENT NOT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 429684 | TRUERESULT | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PS2290 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |