FDA Adverse Event Injury Summary report: N

PENTAX

MDR report key: 4894049 · Received July 2, 2015

Report

Report Number
2518897-2015-00019
Event Type
Injury
Date Received
July 2, 2015
Date of Event
January 1, 2014
Report Date
January 23, 2015
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
FDT
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
MA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DATE RECEIVED BY MFR: 09/27/2016. FOLLOW UP#1. IF FOLLOW-UP, WHAT TYPE?: ADDITIONAL INFORMATION. NO; NOT RETURNED TO MANUFACTURER. LABELED FOR SINGLE USE?: NO. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE FOLLOWING PARAGRAPH WAS REMOVED "BASED ON THE INFORMATION RECEIVED FROM THE COMPLAINANT AND THE INVESTIGATION COMPLETED BY PENTAX MEDICAL ON (B)(6) 2016, IT IS REASONABLE TO SUGGEST USER ERROR AS THE ROOT CAUSE OF THE EVENT, DUE TO THE FACILITY NOT PROPERLY PERFORMING MANUAL CLEANING OF THE DUODENOSCOPE".

Description of Event or Problem · 1

PENTAX MEDICAL CONTACTED THE FACILITY ON 08/10/2015 IN ORDER TO RECEIVE INFORMATION ON PATIENT STATUS, MODEL/SERIAL OF THE DUODENOSCOPE INVOLVED, ADDITIONAL INFORMATION/CLARIFICATION ON THE FACILITY'S ROOT CAUSE ANALYSIS, WHICH BRUSHES ARE USED TO CLEAN THE CYLINDERS OF ENDOSCOPES, AND WHICH DETERGENTS (ENZYMATIC AND NON-ENZYMATIC) AND HIGH LEVEL DISINFECTANTS ARE USED DURING REPROCESSING OF THE ENDOSCOPES. NO FURTHER INFORMATION WAS PROVIDED FROM THE FACILITY IN REGARDS TO THIS EVENT. ON 07/24/2015 INFORMATION WAS RECEIVED FROM THE FACILITY WHICH STATED THE FACILITY BEGAN TESTING DUODENOSCOPES USING THE HEALTH MARK INDUSTRIES CHANNEL CHECK (3-IN-1 RESIDUAL SOIL TEST FOR LUMENED DEVICES) ON (B)(6) 2015. THE FACILITY STATED A RESIDUAL SOIL TEST, WHICH TESTS FOR THE PRESENCE OF CARBOHYDRATES, PROTEIN AND BLOOD IS PERFORMED AFTER EACH WASHING/REPROCESSING OF A DUODENOSCOPE AND PRIOR TO BEING USED IN A PATIENT CARE SETTING, AS A RISK MITIGATION STEP. THE PENTAX REGIONAL SERVICE MANAGER VISITED THE FACILITY DURING NOVEMBER 3-6, 2015 TO OBSERVE THE REPROCESSING TECHNIQUE FOLLOWED AT THE FACILITY. DURING THE VISIT, THE SERVICE MANAGER LEARNED THAT ERCP SCOPES AT THIS FACILITY CAN BE IN USE FOR ONE TO FOUR HOURS DURING A PROCEDURE, AND EVEN THOUGH THEY ARE TAGGED SO THEY GET IMMEDIATE ATTENTION WHEN THEY ARE BROUGHT INTO THE REPROCESSING ROOM, THE BIO-LOAD HAS BEEN IN AND ON THE SCOPES LONG ENOUGH TO ADHERE TO THE CHANNEL WALLS OR ANY OTHER SURFACE OF THE DUODENOSCOPE. DURING THE OBSERVATION OF THE REPROCESSING TECHNIQUE, THE SERVICE MANAGER REALIZED THAT THE REPROCESSING TECHNICIANS WERE NOT USING THE PURGE FUNCTION BUTTON ON THE CUSTOM ULTRASONICS WASH SINKS. THE SERVICE MANAGER STATED THE PURGE FUNCTION, AS PART OF THE MANUAL ENDOSCOPE CLEANING PROCESS, ALLOWS WARM ENZYMATIC SOAP TO FLOW THROUGH ALL OF THE LUMENS (ESPECIALLY THE BIOPSY CHANNEL) OF THE ENDOSCOPES, AND PERFORMING THIS FUNCTION HAS A SIGNIFICANT OUTCOME TOWARD GETTING PATIENT DEBRIS CLEANED FROM THE CHANNEL. AFTER THIS OBSERVATION WAS BROUGHT TO THE FACILITY'S PROCESSING COORDINATOR'S ATTENTION, THE VENDOR FOR CUSTOM ULTRASONICS (AMERICAN DREAM UNLIMITED, LLC) WAS CONTACTED, CONFIRMED WITH THE PROCESSING COORDINATOR THAT THE PURGE FUNCTION MUST BE USED FOR REPROCESSING ALL ENDOSCOPES AND ENSURED THAT ALL SINK COMPONENTS WERE AT FULL CAPACITY AND IN WORKING ORDER. ADDITIONAL INFORMATION WAS RECEIVED FROM THE PROCESSING COORDINATOR ON 11/12/2015 STATING THE REPROCESSING TEAM IS TAKING EXTRA PRECAUTIONARY PROCESSING MEASURES WITH ALL DUODENOSCOPES AS THEY ARE NOW BEING PLACED IN THE STERIS MACHINE, IN ADDITION TO THE POST PROCEDURE PRE-CLEANING, THE MANUAL SINK CLEANING, THE AER CLEANING AND OPA HIGH LEVEL DISINFECTANT CYCLE. IN ADDITION, ALL REPROCESSING TECHNICIANS HAVE BEEN RE-TRAINED/EDUCATED TO ENSURE THE PURGE FUNCTION ON THE CUSTOM ULTRASONICS SINKS IS PERFORMED FOR EVERY ENDOSCOPE AS PART OF THE MANUAL ENDOSCOPE CLEANING PROCESS. ONCE ALL REPROCESSING TECHNICIANS WERE RE-TRAINED/EDUCATED TO ENSURE THE PURGE FUNCTION ON THE CUSTOM ULTRASONICS SINKS IS PERFORMED FOR EVERY ENDOSCOPE AS PART OF THE MANUAL ENDOSCOPE CLEANING PROCESS, NO FURTHER REPORTS OF INFECTION HAVE BEEN RECEIVED FROM THE FACILITY. IN ADDITION, AS A RISK MITIGATION STEP, THE FACILITY BEGAN TESTING DUODENOSCOPES USING THE HEALTH MARK INDUSTRIES CHANNEL CHECK AFTER EACH WASHING/REPROCESSING OF A DUODENOSCOPE AND PRIOR TO BEING USED IN A PATIENT CARE. BASED ON THE INFORMATION RECEIVED FROM THE COMPLAINANT AND THE INVESTIGATION COMPLETED BY PENTAX MEDICAL ON 09/27/2016, IT IS REASONABLE TO SUGGEST USER ERROR AS THE ROOT CAUSE OF THE EVENT, DUE TO THE FACILITY NOT PROPERLY PERFORMING MANUAL CLEANING OF THE DUODENOSCOPE. NO FURTHER INFORMATION HAS BEEN RECEIVED FOR THIS EVENT, THEREFORE, PENTAX MEDICAL CONSIDERS THIS MEDWATCH REPORT CLOSED.

Description of Event or Problem · 1

THE DUODENOSCOPE WAS RECEIVED BY PENTAX MEDICAL ON (B)(6) 2014 FOR EVALUATION. DUE TO THE NATURE OF THE EVENT, ALL PERSONNEL WERE TRAINED ON THE PROTOCOL TO BE FOLLOWED BEFORE THE START OF THE INSPECTION AND DISASSEMBLY. THE DUODENOSCOPE WAS REPROCESSED ACCORDING TO PROCEDURE AND INSPECTED. THE PENTAX SERVICE INSPECTION FINDINGS INCLUDED THE FOLLOWING: UP/DOWN CONTROL KNOB/LEVER CRACKED. FLUID INVASION NOT OBSERVED IN CONTROL BODY. FLUID INVASION NOT OBSERVED IN PVE CONNECTOR. LEAK AT UP/DOWN CONTROL KNOB/LEVER. THE DUODENOSCOPE WAS THEN TRANSPORTED TO A FUME HOOD, WHERE ALL DISASSEMBLY PROCESSES TOOK PLACE. UPON DISASSEMBLY OF THE DUODENOSCOPE, THE FOLLOWING WERE PERFORMED: THE INTERNAL CHANNEL WAS INSPECTED, IN WHICH NO SIGN OF DEBRIS WAS FOUND BEFORE OR AFTER THE CHANNEL WAS CUT OPEN LENGTHWISE. DISTAL END SIDE (DISTAL CAP) AND THE ELEVATOR MECHANISM WERE INSPECTED. A VERY MILD SIGN OF GOUGING AND POSSIBLE SHREDDING OF THE MATERIAL AT THE TOP SECTION OF THE DISTAL CAP WAS VISUALIZED. NO OTHER VISIBLE SIGNS OF DEBRIS ANYWHERE AROUND THE DISTAL CAP AREA HOWEVER, DEBRIS WAS VISUALIZED IN THE DISTAL CAP. AS THE ELEVATOR MECHANISM SECTION WAS BEING REMOVED, SEVERAL SIGNS OF WHITE STAINS WERE VISUALIZED ON BOTH SIDES OF THE DISTAL BODY WHICH HOUSES THE ELEVATOR BODY. DURING DISASSEMBLY OF THE DISTAL BODY SIDE COVER FOR THE DEFLECTOR SIDE, SEVERAL BROWNISH LOOKING DEBRIS/DUST PARTICLES WERE VISUALIZED IN THE SECTION WHERE THE DEFLECTOR BODY LINK IS INSTALLED. WHEN THE DEFLECTOR BODY LINK (ELEVATOR PISTON WITH TWO O-RINGS) WAS REMOVED AND SEPARATED FROM THE DISTAL BODY, BROWNISH/RUST LOOKING DEBRIS WAS VISUALIZED AT THE END OF THE DEFLECTOR BODY LINK. THE SAME BROWNISH/RUST LOOKING DEBRIS WAS VISUALIZED AT THE RECESS SECTION OF THE DISTAL BODY. THE DEFLECTOR BODY LINK O-RINGS WERE INTACT AND HAD NO SIGNS OF DEBRIS. THE INSPECTORS PAID SPECIAL ATTENTION FOR ANY COMPROMISED COMPONENTS DURING THE DISASSEMBLY PROCESS AND NONE WERE FOUND. THE DISASSEMBLED PARTS WERE SHIPPED TO (B)(6) ON (B)(6) 2015 TO PARTAKE IN A STUDY. A DEVICE HISTORY REVIEW WAS PERFORMED ON (B)(6) 2014 CONFIRMING THE VIDEO DUODENOSCOPE WAS MANUFACTURED UNDER NORMAL CONDITIONS, PASSED ALL REQUIRED INSPECTIONS, AND WAS RELEASED ACCORDINGLY. ALSO, THERE WERE NO REWORKS OR CONCESSIONS AND THE DATES OF APPROVAL FOR SHIPMENT AND ACTUAL DATE SHIPPED WERE CONFIRMED. PENTAX MEDICAL CONTACTED THE FACILITY ON (B)(6) 2015 AND (B)(6) 2015 FOR ADDITIONAL INFORMATION ON THE FACILITY'S ROOT CAUSE ANALYSIS, REPROCESSING AGENTS (I.E. DETERGENTS (ENZYMATIC AND NON-ENZYMATIC), HIGH LEVEL DISINFECTANTS, AND AUTOMATED ENDOSCOPE REPROCESSORS), AND REPROCESSING TECHNICIAN TRAINING. A RESPONSE WAS RECEIVED FROM THE FACILITY ON (B)(6) 2015 STATING THE FIRST PHASE OF THEIR ROOT CAUSE ANALYSIS PROVIDED SOME GENERAL BUT IMPORTANT OBSERVATIONS. IT SHOWED THAT EXPOSURE TO A CERTAIN DUODENOSCOPE WAS ASSOCIATED WITH FUTURE INFECTION WITH SPECIFIC STRAINS OF (B)(6). THE FACILITY STATED THEY WERE UNABLE TO DETERMINE WHERE ON THE DUODENOSCOPE THE NIDUS FOR BACTERIA MIGHT LAY OR THE SPECIFIC CAUSE OF THE TRANSMISSION. IN ADDITION, NO DEFECT IN THE DUODENOSCOPE CONSTRUCTION AND FUNCTION OR BREAK IN THE DUODENOSCOPE REPROCESSING WAS IDENTIFIED. THE FACILITY ALSO STATED THEY ARE DESIGNING A HEIGHTENED MONITORING SYSTEM AND PERHAPS FUTURE OBSERVATIONS WILL ALLOW MORE SPECIFIC DEDUCTIONS. NO INFORMATION WAS RECEIVED FROM THE FACILITY IN REGARDS TO WHAT DETERGENTS (ENZYMATIC AND NON-ENZYMATIC), AND HIGH LEVEL DISINFECTANTS ARE USED DURING REPROCESSING OF THE DUODENOSCOPE. INFORMATION REGARDING AUTOMATED ENDOSCOPE REPROCESSORS WAS RECEIVED. THE FACILITY DID STATE THAT THEIR TECHNICIANS ARE TRAINED AT THE BEGINNING OF THEIR EMPLOYMENT, UPDATED AT LEAST YEARLY BY MASSACHUSETTS GENERAL HOSPITAL EXPERTS, PENTAX AND OTHER MANUFACTURES' REPRESENTATIVES AND THEIR PERFORMANCE IS REGULARLY OBSERVED BY ONSITE EXPERT SUPERVISORS. PENTAX MEDICAL CONTACTED THE FACILITY ON (B)(6) 2015 AND (B)(6) 2015 IN ORDER TO RECEIVE ADDITIONAL INFORMATION/CLARIFICATION ON THE FACILITY'S ROOT CAUSE ANALYSIS, WHICH BRUSHES ARE USED TO CLEAN THE CYLINDERS OF ENDOSCOPES, AND WHICH DETERGENTS (ENZYMATIC AND NON-ENZYMATIC) AND HIGH LEVEL DISINFECTANTS ARE USED DURING REPROCESSING OF THE ENDOSCOPES. NO ADDITIONAL INFORMATION/CLARIFICATION WAS RECEIVED. PENTAX MEDICAL CONTACTED THE FACILITY ON (B)(6) 2015, (B)(6) 2015 AND (B)(6) 2015 TO INQUIRE IF THE REPORTED DEATH WAS A RESULT OF THE (B)(6) BACTEREMIAS, AND TO RETRIEVE ALL OTHER PENDING INFORMATION. NO FURTHER INFORMATION WAS PROVIDED FROM THE FACILITY IN REGARDS TO THIS EVENT. ON (B)(6) 2015 INFORMATION WAS RECEIVED FROM THE FACILITY WHICH STATED THE FACILITY BEGAN TESTING DUODENOSCOPES USING THE HEALTH MARK INDUSTRIES CHANNEL CHECK (3-IN-1 RESIDUAL SOIL TEST FOR LUMENED DEVICES) ON (B)(6) 2015. THE FACILITY STATED A RESIDUAL SOIL TEST, WHICH TESTS FOR THE PRESENCE OF CARBOHYDRATES, PROTEIN AND BLOOD IS PERFORMED AFTER EACH WASHING/REPROCESSING OF A DUODENOSCOPE AND PRIOR TO BEING USED IN A PATIENT CARE SETTING, AS A RISK MITIGATION STEP. THE PENTAX REGIONAL SERVICE MANAGER VISITED THE FACILITY DURING (B)(6) 2015 TO OBSERVE THE REPROCESSING TECHNIQUE FOLLOWED AT THE FACILITY. DURING THE VISIT, THE SERVICE MANAGER LEARNED THAT ERCP SCOPES AT THIS FACILITY CAN BE IN USE FOR ONE TO FOUR HOURS DURING A PROCEDURE, AND EVEN THOUGH THEY ARE TAGGED SO THEY GET IMMEDIATE ATTENTION WHEN THEY ARE BROUGHT INTO THE REPROCESSING ROOM, THE BIO-LOAD HAS BEEN IN AND ON THE SCOPES LONG ENOUGH TO ADHERE TO THE CHANNEL WALLS OR ANY OTHER SURFACE OF THE DUODENOSCOPE. DURING THE OBSERVATION OF THE REPROCESSING TECHNIQUE, THE SERVICE MANAGER REALIZED THAT THE REPROCESSING TECHNICIANS WERE NOT USING THE PURGE FUNCTION BUTTON ON THE CUSTOM ULTRASONICS WASH SINKS. THE SERVICE MANAGER STATED THE PURGE FUNCTION, AS PART OF THE MANUAL ENDOSCOPE CLEANING PROCESS, ALLOWS WARM ENZYMATIC SOAP TO FLOW THROUGH ALL OF THE LUMENS (ESPECIALLY THE BIOPSY CHANNEL) OF THE ENDOSCOPES, AND PERFORMING THIS FUNCTION HAS A SIGNIFICANT OUTCOME TOWARD GETTING PATIENT DEBRIS CLEANED FROM THE CHANNEL. AFTER THIS OBSERVATION WAS BROUGHT TO THE FACILITY'S PROCESSING COORDINATOR'S ATTENTION, THE VENDOR FOR CUSTOM ULTRASONICS ((B)(6)) WAS CONTACTED, CONFIRMED WITH THE PROCESSING COORDINATOR THAT THE PURGE FUNCTION MUST BE USED FOR REPROCESSING ALL ENDOSCOPES AND ENSURED THAT ALL SINK COMPONENTS WERE AT FULL CAPACITY AND IN WORKING ORDER. ADDITIONAL INFORMATION WAS RECEIVED FROM THE PROCESSING COORDINATOR ON (B)(6) 2015 STATING THE REPROCESSING TEAM IS TAKING EXTRA PRECAUTIONARY PROCESSING MEASURES WITH ALL DUODENOSCOPES AS THEY ARE NOW BEING PLACED IN THE STERIS MACHINE, IN ADDITION TO THE POST PROCEDURE PRE-CLEANING, THE MANUAL SINK CLEANING, THE AER CLEANING AND OPA HIGH LEVEL DISINFECTANT CYCLE. IN ADDITION, ALL REPROCESSING TECHNICIANS HAVE BEEN RE-TRAINED/EDUCATED TO ENSURE THE PURGE FUNCTION ON THE CUSTOM ULTRASONICS SINKS IS PERFORMED FOR EVERY ENDOSCOPE AS PART OF THE MANUAL ENDOSCOPE CLEANING PROCESS. ONCE ALL REPROCESSING TECHNICIANS WERE RE-TRAINED/EDUCATED TO ENSURE THE PURGE FUNCTION ON THE CUSTOM ULTRASONICS SINKS IS PERFORMED FOR EVERY ENDOSCOPE AS PART OF THE MANUAL ENDOSCOPE CLEANING PROCESS, NO FURTHER REPORTS OF INFECTION HAVE BEEN RECEIVED FROM THE FACILITY. IN ADDITION, AS A RISK MITIGATION STEP, THE FACILITY BEGAN TESTING DUODENOSCOPES USING THE HEALTH MARK INDUSTRIES CHANNEL CHECK AFTER EACH WASHING/REPROCESSING OF A DUODENOSCOPE AND PRIOR TO BEING USED IN A PATIENT CARE. NO FURTHER INFORMATION HAS BEEN RECEIVED FOR THIS EVENT, THEREFORE, PENTAX MEDICAL CONSIDERS THIS MEDWATCH REPORT CLOSED.

Description of Event or Problem · 1

PENTAX MEDICAL RECEIVED A REPORT ON (B)(6) 2015 INDICATING 1 PATIENT TESTED POSITIVE FOR A SINGLE STRAIN TYPE OF CEFTRIAXONE-RESISTANT E. COLI POST-ERCP PROCEDURE. NO INFORMATION ON THE MEDICAL DEVICE USED DURING THE ERCP PROCEDURE WAS RECEIVED AT THE TIME OF THIS REPORT. PATIENT (B)(6) MET THE CASE OF "IMMEDIATE BACTEREMIA POST-ERCP" MEANING PATIENT HAD (+) BLOOD CULTURE POST-ERCP WITHIN 72 HOURS. THE FACILITY PROTOCOL FOR INFECTIONS THAT ARE PART OF THE HOSPITAL SURVEILLANCE PLAN, WHICH INCLUDES BACTEREMIAS, IS THAT THE DEPARTMENT LEADERSHIP AND THE PROVIDER ARE NOTIFIED OF THE INFECTION; ANY TRENDS IDENTIFIED AND A REVIEW OF THE INFECTION (ROOT CAUSE ANALYSIS) IS COMPLETED. THE FACILITY ALSO INDICATED THAT PATIENTS UNDERGOING ERCP ARE BEING TREATED FOR A RANGE OF PANCREATIC AND BILIARY DISEASES; IT WOULD BE DIFFICULT TO DETERMINE SPECIFIC SIGNS AND SYMPTOMS OTHER THAN FEVER/CHILLS AND MALAISE THAT WOULD BE RELATED TO THE POST-PROCEDURE BACTEREMIAS. PATIENT WAS TREATED FOR BACTEREMIAS WITH APPROPRIATE ANTIBIOTICS. PATIENT WAS NOT RECALLED FOR FURTHER SCREENING AS THE MARKER FOR THIS OUTBREAK INVESTIGATION WAS A (+) BLOOD CULTURE FOR CEFTRIAXONE-RESISTANT E.COLI; ALL (+) BLOOD CULTURES ARE REVIEWED DAILY BY INFECTION PREVENTIONISTS. PENTAX MEDICAL CONTACTED THE INITIAL REPORTER VIA EMAIL ON (B)(6) 2015, (B)(6) 2015, (B)(6) 2015 AND (B)(6) 2015 TO CONFIRM THE MEDICAL DEVICE USED DURING THE ERCP PROCEDURE. ALSO, THE INITIAL REPORTER WAS CONTACTED VIA EMAIL ON (B)(6) 2015 AND (B)(6) 2015 IN REGARDS TO CURRENT PATIENT STATUS. NO INFORMATION ON THE MEDICAL DEVICE INVOLVED IN THIS EVENT OR CURRENT PATIENT STATUS INFORMATION FOR PATIENT ID (B)(6) HAVE BEEN RECEIVED FROM THE FACILITY TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
433163 PENTAX VIDEO DUODENOSCOPE FDT HOYA CORPORATION PENTAX TOKYO OFFICE

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention