LAS VEGAS PLUS BLUE ALUMINUM HANDLE CANNULA
Report
- Report Number
- 1718903-2015-00001
- Event Type
- Malfunction
- Date Received
- July 1, 2015
- Date of Event
- May 1, 2015
- Report Date
- June 22, 2015
- Manufacturer
- SHIPPERT MEDICAL TECHNOLOGIES CORP.
- Product Code
- GEA
- PMA / PMN Number
- K022852
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NURSE
Narratives
THE DEVICE WAS SUBJECTED TO EXTERNAL FORCE TO CAUSE IT TO BREAK. THE MATERIAL OF CONSTRUCTION, STAINLESS STEEL, IS NOT ONLY COMMON TO CANNULAS, BUT IS VERY STRONG. THE CANNULA HAD TO BE FLEXED REPEATEDLY TO BREAK. MAUDE DATABASE WAS EXAMINED FINDING 41 ENTRIES ASSOCIATED WITH "BREAK". NONE INVOLVED A CANNULA. REPEATED ATTEMPTS WERE MADE TO COLLECT MISSING DATA. NURSE AND DOCTOR DO NOT RESPOND TO ALL QUESTIONS AND ARE NOT TIMELY IN RESPONSES.
PATIENT WENT TO ER AFTER ABDOMINOPLASTY WITH LIPOSUCTION BECAUSE OF DIZZINESS. AFTER EVALUATION, ER FOUND BROKEN PIECE OF LIPOSUCTION CANNULA INSIDE THE PATIENT NEAR SURGICAL SITE. ER REPORTED RESULT TO SURGEON WHO REPORTED BROKEN CANNULA TO US. SURGEON DETERMINED THAT THE BROKEN PIECE COULD REMAIN IN THE PATIENT UNTIL PARTIAL HEALING TAKES PLACE (THEREFORE THIS OBSERVATION IS NOT SERIOUS AND WE DO NOT BELIEVE IT IS AN ADVERSE EVENT) AND THE SURGEON WOULD REMOVE A PIECE OF CANNULA IN THE (B)(6) 2015, TIME FRAME. THE CANNULA WAS A RE-USABLE CANNULA THAT HAD BEEN USED AT LEAST FOUR TIMES PRIOR TO THIS BREAK. (B)(4) HAS BEEN DISTRIBUTING THIS TYPE OF CANNULA FOR AT LEAST EIGHT YEARS (OVER (B)(4) CANNULAS) WITHOUT A PROBLEM OF THIS TYPE. AFTER EVALUATION OF THE PIECE THE DOCTOR STILL HAD AND SENT TO US FOR EVALUATION, (B)(4) BELIEVES THE CANNULA BROKE DUE TO AN EVENT DURING THE SURGERY, NOT A WEAKNESS IN THE CANNULA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 429354 | LAS VEGAS PLUS BLUE ALUMINUM HANDLE CANNULA | CANNULA, SURGICAL | GEA | SHIPPERT MEDICAL TECHNOLOGIES CORP. | 3-H-BA-LV+-5X30 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| O |