FDA Adverse Event
Malfunction
Summary report: N
PEGASYS
MDR report key: 489082
·
Received September 25, 2003
Report
- Report Number
- MW1029682
- Event Type
- Malfunction
- Date Received
- September 25, 2003
- Date of Event
- September 19, 2003
- Report Date
- September 25, 2003
- Manufacturer
- ADAC LABORATORIES
- Product Code
- KPW
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- KY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
COLLIMATOR EXCHANGE CARRIER WAS OUT OF ADJUSTMENT. FELL ON EMPLOYEE CAUSING STRAIN TO BACK AND INJURY TO HIP AND RIGHT FOOT. FIELD REP NOTIFIED. ON-SITE VISIT CONDUCTED OF EQUIPMENT. EXCHANGE CARRIAGE WAS OUT OF ADJUSTMENT. ADJUSTED CARRIER AND CLEANED BEARINGS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PEGASYS | COLLIMATOR EXCHANGER CARRIER | KPW | ADAC LABORATORIES | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR | Other |