Description of Event or Problem · 1

COLLIMATOR EXCHANGE CARRIER WAS OUT OF ADJUSTMENT. FELL ON EMPLOYEE CAUSING STRAIN TO BACK AND INJURY TO HIP AND RIGHT FOOT. FIELD REP NOTIFIED. ON-SITE VISIT CONDUCTED OF EQUIPMENT. EXCHANGE CARRIAGE WAS OUT OF ADJUSTMENT. ADJUSTED CARRIER AND CLEANED BEARINGS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PEGASYS COLLIMATOR EXCHANGER CARRIER KPW ADAC LABORATORIES * *

Patients

Seq Age Sex Outcome Treatment
1 32 YR Other