GREENLIGHT MOXY FIBER OPTIC
Report
- Report Number
- 2937094-2015-00688
- Event Type
- Malfunction
- Date Received
- July 5, 2015
- Date of Event
- June 23, 2015
- Report Date
- June 26, 2015
- Manufacturer
- AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
- Product Code
- GEX
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4).
(B)(4). PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER (B)(4): THE FIBER/GLASS CAP SHOWS A CIRCUMFERENTIAL FRACTURE AT DISTAL SIDE OF FIBER/CAP FUSION ZONE AT THE BEVEL EDGE; THE GLASS CAP EXHIBITS VERY MILD DEVITRIFICATION AT THE OUTPUT WINDOW; THE METAL CAP EXHIBITS VERY MILD DETRITUS ADHESION. BASED ON DEVICE ANALYSIS, THE POTENTIAL FOR FORWARD FIRING MAY EXIST. PROBABLE ROOT CAUSE: BASED ON THE DEVICE ANALYSIS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH WOULD LIMIT THE PERFORMANCE OF THE FIBER.
THE EVENT OCCURRED AT 183,602 JOULES AND 27:09 MINUTES OF USAGE.
IT WAS REPORTED THAT DURING A BPH PROSTATE PROCEDURE "FORWARD FIRING" WAS NOTED. THE FIBER WAS EXCHANGED AND THE PROCEDURE COMPLETED. PATIENT OUTCOME: ¿NO INJURY TO PATIENT " WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 434245 | GREENLIGHT MOXY FIBER OPTIC | POWERED LASER SURGICAL INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) | 0010-2400 | 509A |