ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2015-22165
- Event Type
- Malfunction
- Date Received
- July 5, 2015
- Report Date
- June 20, 2015
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1: DATE OF SUBMISSION 07/21/2015. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/07/2015 WITH THE FOLLOWING FINDINGS: THE UP, DOWN, AND OK BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. CONTAMINATION WAS NOT FOUND ON THE BUTTON CONTACTS, BUT MOISTURE DAMAGE WAS FOUND TO THE KEYPAD CONNECTOR. UNRELATED TO THE INITIAL ALLEGATION, THE DISPLAY SCREEN WAS DIM A ND DISCOLORED. THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS AT THE TIME OF THIS REPORT. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2015, A REPORTER CONTACTED ANIMAS ALLEGING THAT THE UP, DOWN, AND OK BUTTONS WERE UNDERRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED THROUGH TROUBLESHOOTING. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 434244 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |