FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4890396 · Received July 5, 2015

Report

Report Number
2531779-2015-22165
Event Type
Malfunction
Date Received
July 5, 2015
Report Date
June 20, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 07/21/2015. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/07/2015 WITH THE FOLLOWING FINDINGS: THE UP, DOWN, AND OK BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. CONTAMINATION WAS NOT FOUND ON THE BUTTON CONTACTS, BUT MOISTURE DAMAGE WAS FOUND TO THE KEYPAD CONNECTOR. UNRELATED TO THE INITIAL ALLEGATION, THE DISPLAY SCREEN WAS DIM A ND DISCOLORED. THE BATTERY COMPARTMENT WAS FOUND TO BE CRACKED.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS AT THE TIME OF THIS REPORT. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2015, A REPORTER CONTACTED ANIMAS ALLEGING THAT THE UP, DOWN, AND OK BUTTONS WERE UNDERRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED THROUGH TROUBLESHOOTING. THERE WAS NO ALLEGATION OF AN ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
434244 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 11 YR