FDA Adverse Event Malfunction Summary report: N

I-STAT CHEM8+ CARTRIDGE

MDR report key: 4890349 · Received July 5, 2015

Report

Report Number
2245578-2015-00034
Event Type
Malfunction
Date Received
July 5, 2015
Date of Event
June 14, 2015
Report Date
June 15, 2015
Manufacturer
ABBOTT POINT OF CARE
Product Code
JGS
PMA / PMN Number
K053110
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE INVESTIGATION WAS COMPLETED ON 07/17/2015. RETAIN PRODUCT WAS TESTED AND IS FUNCTIONING ACCORDING TO SPECIFICATION. RETURN PRODUCT WAS NOT AVAILABLE FOR INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

ON (B)(6) 2015, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT HEMATOCRIT RESULT ON A PATIENT. THERE WAS NO PATIENT INFORMATION AT THE TIME OF THIS REPORT. CUSTOMER IS NOT RETURNING PRODUCT FOR INVESTIGATION. DATE METHOD SAMPLE RESULT TIME HCT (%) (B)(6) 2015 I-STAT A 12:25 <10% (B)(6) 2015 I-STAT A 12:41 35% (B)(6) 2015 LAB B 08:00 28% THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
434309 I-STAT CHEM8+ CARTRIDGE CHEM8+ CARTRIDGE JGS ABBOTT POINT OF CARE NA H15053

Patients

Seq Age Sex Outcome Treatment
1