FDA Adverse Event Malfunction Summary report: N

DIAGNODENT CLASSIC

MDR report key: 4889794 · Received July 3, 2015

Report

Report Number
3003637274-2015-00013
Event Type
Malfunction
Date Received
July 3, 2015
Date of Event
June 4, 2015
Report Date
June 5, 2015
Manufacturer
KAVO DENTAL GMBH
Product Code
NBL
PMA / PMN Number
K983658
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

THE EVALUATION PERFORMED BY AN INDEPENDENT INSTITUTE SHOWED THAT THE ORIGIN OF THE HEAT MUST BE FROM A SINGLE BATTERY CELL WITHIN THE BATTERY PACK WHICH GOT TOO HOT. ACCORDING TO THE INSTITUTE IT IS NOT POSSIBLE TO IDENTIFY THE ROOT CAUSE FOR THE OVERHEATING OF A SINGLE CELL AS THERE ARE VARIOUS POSSIBILITIES LIKE: FAILURE IN THE CELL DUE TO A HIT, E.G. FROM DROPPING, INACCURATE WELDING POINT ON ONE POLE AND THEREFORE A HIGHER CONTACT RESISTANCE, EXHAUSTIVE DISCHARGE OF A SINGLE CELL AND FURTHER POSSIBILITIES. BESIDE THE ISSUE WE REPORTED WITH MDR 3003637274-2015-00005 (IN THIS CASE IT WAS A DEFECT DUE TO DROPPING OF THE BATTERY PACK) WE ARE NOT AWARE OF FURTHER BURNING ISSUES. AS THERE ARE MORE THAN (B)(4) BATTERY PACKS DELIVERED AND ONLY (B)(4) INCIDENTS ARE KNOWN IT IS VERY LIKELY THAT ALSO THIS SECOND INCIDENT IS CAUSED BY A HANDLING FAILURE. THEREFORE THE CONCLUSION CODE IS SELECTED CORRESPONDINGLY. THE USER INSTRUCTION CONTAINS A NOTE THAT RECHARGEABLE BATTERIES MUST NOT BE DROPPED OR RECEIVED HITS FROM OTHER SOURCES.

Additional Manufacturer Narrative · 1

AS IT IS NOT POSSIBLE TO SEND THE CHARGEABLE BATTERY PACK BACK TO THE VENDOR IN (B)(4) FOR EVALUATION WE CONSULTED AN INDEPENDENT INSTITUTE TO ANALYZE SOME CHARGEABLE BATTERY PACK OUT OF THE SAME PRODUCTION PERIOD TO VERIFY THAT THEY MEET SPECIFICATION. THE PROCESS IS STARTED AND TAKES AT LEAST ADDITIONAL 3 WEEKS. AS SOON AS WE HAVE A RESULT WE WILL ISSUE A FOLLOW UP REPORT.

Description of Event or Problem · 1

ACCORDING TO INFORMATION FROM DENTAL OFFICE THE BATTERY PACK WAS GETTING HOT WHILE IN USE ON THE DIAGNODENT CARIES DETECTOR. DOCTOR REMOVED THE BATTERY PACK FROM THE DIAGNODENT, THEN IT STARTED TO SPARK AND CAUGHT FIRE. NOBODY WAS INJURED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
433341 DIAGNODENT CLASSIC CARIES DETECTOR NBL KAVO DENTAL GMBH KAVO DIAGNODENT 2095

Patients

Seq Age Sex Outcome Treatment
1