FDA Adverse Event Malfunction Summary report: N

DIMENSION VISTA® SYSTEM

MDR report key: 4889370 · Received July 2, 2015

Report

Report Number
2517506-2015-00155
Event Type
Malfunction
Date Received
July 2, 2015
Date of Event
June 26, 2014
Report Date
July 30, 2015
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
Product Code
CGX
PMA / PMN Number
K061238
Removal / Correction Number
2517506-07152015-003-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
MEDICAL TECHNOLOGIST

Narratives

Additional Manufacturer Narrative · 1

ORIGINAL MDR SUBMITTED 2015-07-02. AN URGENT MEDICAL DEVICE CORRECTION (US) AND AN URGENT FIELD SAFETY NOTICE (OUS) DATED JULY 2015 WAS PROVIDED TO ALL CUSTOMERS WHO RECEIVED THE AFFECTED LOTS TO NOTIFY THEM OF THE FOLLOWING: SIEMENS HAS CONFIRMED THAT THE DIMENSION CREA AND DIMENSION VISTA CREA MAY PRODUCE FALSELY DEPRESSED RESULTS UP TO -12.7% AND -22% RESPECTIVELY, FOR A CREATININE CONCENTRATION OF 1.5 MG/DL WHEN HEMOLYSIS EXCEEDS 200 MG/DL AND 300 MG/DL RESPECTIVELY. CUSTOMERS WERE INSTRUCTED TO REVIEW THE LETTER WITH THEIR MEDICAL DIRECTOR. CUSTOMERS WERE INSTRUCTED TO REVIEW AND UPDATE THEIR ESTABLISHED PROCEDURES FOR REPORTING CREA RESULTS FOR HEMOLYZED SAMPLES BASED ON THE INFORMATION IN THE LETTER. CUSTOMERS WERE INSTRUCTED TO UPDATE THE H INDEX FOR HEMOLYSIS IF THAT FEATURE IS BEING USED ON THEIR SYSTEM. (B)(4).

Additional Manufacturer Narrative · 1

SIEMENS HEALTHCARE DIAGNOSTICS HAS CONDUCTED INVESTIGATIONS ON HEMOGLOBIN INTERFERENCE WITH THE DIMENSION VISTA® K1033 CREATININE (CREA) FLEX REAGENT CARTRIDGES. INVESTIGATIONS HAVE DETERMINED THAT DIMENSION VISTA® CREATININE (CREA) MAY PRODUCE DEPRESSED RESULTS UP TO 22% (OR APPROXIMATELY 0.3 MG/DL [27 ¿MOL/L]) FOR A CREATININE RESULT AT 1.5 MG/DL [133 ¿MOL/L] WHEN HEMOLYSIS EXCEEDS 200 MG/DL [0.12 MMOL/L] OF HEMOGLOBIN. PERCENT HEMOGLOBIN INTERFERENCE WITH DIMENSION VISTA® CREA IS DEPENDENT ON THE CREATININE CONCENTRATION; AT A CREATININE CONCENTRATION OF 5.0 MG/DL [442 ¿MOL/L], HEMOGLOBIN INTERFERENCE AT CONCENTRATIONS UP TO A 1000 MG/DL [0.62 MMOL/L] IS < 10%. ALTERNATELY, WITH AN APPROXIMATE BIAS OF 0.3 MG/DL [27 ¿MOL/L]), THE PERCENT HEMOGLOBIN INTERFERENCE MAY BE HIGHER AT LOW CREATININE CONCENTRATIONS (< 0.8 MG/DL [< 71 ¿MOL/L]). DEPRESSED CREATININE RESULTS DUE TO HEMOLYSIS AS DESCRIBED ABOVE MAY LEAD TO INCREASED ESTIMATED GLOMERULAR FILTRATION RATE (EGFR) VALUES AND/OR POTENTIAL MISINTERPRETATION OF ACUTE KIDNEY INJURY (AKI). THE POTENTIAL FOR INJURY IS REMOTE DUE TO CONTINUED MONITORING, ADDITIONAL DIAGNOSTIC TESTING AND CORRELATION TO PATIENT HISTORY AND PRESENTATION. DEPRESSED CREATININE RESULTS DUE TO HEMOLYSIS AS DESCRIBED ABOVE WOULD NOT BE EXPECTED TO SIGNIFICANTLY IMPACT MEDICAL DECISIONS WHEN USING THIS ASSAY. HOWEVER, SIEMENS HEALTHCARE DIAGNOSTICS IS ISSUING AN URGENT MEDICAL DEVICE CORRECTION (US) AND AN URGENT FIELD SAFETY NOTICE (OUS) DATED (B)(6)2015 ALERTING CUSTOMERS THAT THEY SHOULD REVIEW THEIR CURRENTLY ESTABLISHED PROCEDURE FOR REPORTING CREA RESULTS FOR HEMOLYZED SAMPLES AND UPDATE PROCEDURES AS NECESSARY FOR THE REVISED INTERFERENCE INFORMATION DESCRIBED IN THE COMMUNICATION.

Description of Event or Problem · 1

AN ACCOUNT COMPLAINED OF FALSELY DEPRESSED CREATININE (CREA) RESULTS ON HEMOLYZED PATIENT SAMPLES. IT IS UNKNOWN IF THE RESULTS WERE REPORTED TO THE PHYSICIAN. THE ACCOUNT RAN THE SAMPLES WITH ALTERNATE METHODOLOGIES AND OBTAINED HIGHER RESULTS MORE IN ALIGNMENT WITH PHYSICIAN EXPECTATIONS. IT IS UNKNOWN IF PATIENT TREATMENT WAS ALTERED OR PRESCRIBED ON THE BASIS OF THE DEPRESSED CREA RESULTS. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE DEPRESSED CREA RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
430843 DIMENSION VISTA® SYSTEM CREATININE FLEX® REAGENT CARTRIDGE CGX SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW NOT PROVIDED

Patients

Seq Age Sex Outcome Treatment
1