FDA Adverse Event Injury Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 4888456 · Received June 30, 2015

Report

Report Number
3008642652-2015-04097
Event Type
Injury
Date Received
June 30, 2015
Date of Event
June 2, 2015
Report Date
June 25, 2015
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO DEATH OR DEVICE MALFUNCTION ASSOCIATED WITH THE INAPPROPRIATE DEFIBRILLATION. CODE (OTHER): DEVICE EVALUATION WAS ACCOMPLISHED THROUGH A REVIEW OF THE PT'S DOWNLOADED DATA FILE. REVIEW OF THE DATE DOES NOT INDICATE ANY DEVICE MALFUNCTION RELATED TO THE DEFIBRILLATION EVENT. USAGE OF DEVICE: MONITOR (B)(4) 04/2015 - INITIAL USE. ELECTRODE BELT (B)(4) 07/1011 - REUSE. ADDITIONAL INAPPROPRIATE DEFIBRILLATION NARRATIVE: INAPPROPRIATE DEFIBRILLATION ARE AN ANTICIPATED RISK ASSOCIATED WITH THE USE OF THE LIFEVEST. PATIENTS ARE INSTRUCTED THROUGH ALARMS, VOICE MESSAGES, IFU, AND TRAINING TO PRESS THE RESPONSE BUTTONS TO PREVENT AN INAPPROPRIATE DEFIBRILLATION. THE CURRENT COMMERCIAL INAPPROPRIATE DEFIBRILLATION RATE IS CONSISTENT WITH THE OBSERVED RATE DURING (B)(4). A SUMMARY OF THE SAFETY AND EFFECTIVENESS DATA (SSED), INCLUDING THE INAPPROPRIATE DEFIBRILLATION SAFETY OBJECTIVE SUPPORTING FDA'S APPROVAL OF THE LIFEVEST, CAN BE FOUND AT HTTP://WWW.ACCESSDATA.FDA.GOV/CDRH_DOCS/PDF/P010030B.PDF.

Description of Event or Problem · 1

A UNITED STATES DISTRIBUTOR CONTACTED ZOLL TO REPORT THAT A (B)(6) PT EXPERIENCED AN INAPPROPRIATE DEFIBRILLATION EVENT. RAPID ATRIAL FIBRILLATION CONTRIBUTED TO THE FALSE DETECTION. THE RESPONSE BUTTONS WERE NOT PRESSED DURING THE EVENT. THE PT CONTINUED USE OF THE LIFEVEST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
421058 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 64 YR Hospitalization| O