FDA Adverse Event Malfunction Summary report: N

CUSA EXCEL

MDR report key: 488779 · Received September 23, 2003

Report

Report Number
MW1029632
Event Type
Malfunction
Date Received
September 23, 2003
Date of Event
June 21, 2003
Report Date
September 16, 2003
Manufacturer
VALLEYLAB DIV TYCO HEALTHCARE GR
Product Code
LBK
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
MI, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

ADD'L INFO REC'D FROM MFR 11/11/03: THE MEDWATCH SENT BY THE CUSTOMER IDENTIFIED THE CONSOLE, RATHER THAN THE CUSA TIP. THE CUSA TIP WAS NOT RETURNED TO VALLEYLAB FOR EVAL; HOWEVER, THE HANDPIECE HAD BEEN RETURNED TO VALLEYLAB'S SERVICE DEPT ON JULY 18TH, 2003. THE TEST RESULTS FOUND THE HANDPIECE TO OPERATE PROPERLY AND WITHIN ALL SPECS WHEN TESTED WITH A SPARE TEST TIP. THE HANDPIECE WAS DISASSEMBLED AND NO CONDITIONS WERE FOUND THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE TIP BREAKING. ALTHOUGH THE CAUSE OF THE CUSTOMER'S REPORT CANNOT BE DETERMINED, TIP BREAKAGE IS KNOWN TO OCCUR WHEN THE TIP IS PLACED IN CLOSE PROXIMITY TO ELECTROSURGICAL ENERGY DURING USE. THE TIP WILL PRODUCE A BLUISH COLOR, BUT AGAIN, THE TIP WAS NOT RETURNED FOR EXAM. PRODUCT LABELING CAUTIONS THE USER NOT TO USE THE TIP IN CLOSE PROXIMITY TO ELECTROSURGERY.

Description of Event or Problem · 1

TIP OF THE CUSA (ULTRASOUNIC SURGICAL ASPIRATOR) HANDPIECE BROKE WHILE IN USE. TIP RETRIEVED, NO INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CUSA EXCEL ULTRASONIC SURGICAL ASPIRATOR LBK VALLEYLAB DIV TYCO HEALTHCARE GR CUSAEXCEL NA

Patients

Seq Age Sex Outcome Treatment
1 *