FDA Adverse Event Malfunction Summary report: N

MINIMAL ACCESS ATTACHMENT

MDR report key: 4884747 · Received July 1, 2015

Report

Report Number
1045834-2015-11431
Event Type
Malfunction
Date Received
July 1, 2015
Date of Event
June 2, 2015
Report Date
June 3, 2015
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBE
PMA / PMN Number
PK042783
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE DEVICE HAD TEMPERATURE OVER SPECIFICATION AND THE BEARINGS WERE WORN OUT. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO WORN OUT BEARINGS FROM USE AND SERVICING OVER TIME. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING, IT WAS OBSERVED THAT THE ATTACHMENT DEVICE GOT HOT. THE EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, ADVERSE EVENTS OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
428690 MINIMAL ACCESS ATTACHMENT DRILLS, BURRS, TREPHINES & ACCESSORIES - ATTACHMENT HBE DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1