FDA Adverse Event Malfunction Summary report: N

COGNIS

MDR report key: 4884745 · Received July 1, 2015

Report

Report Number
2124215-2015-07418
Event Type
Malfunction
Date Received
July 1, 2015
Date of Event
May 28, 2015
Report Date
June 24, 2015
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Removal / Correction Number
Z-0087-2015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE REMAINS IN SERVICE. THIS REPORT WILL BE UPDATED SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED UPON RECEIPT AT OUR QUALITY ASSURANCE LABORATORY. REVIEW OF THE DEVICE MEMORY INDICATED THAT A LOW VOLTAGE ALERT, CODE 1003, WAS RECORDED. THE BATTERY VOLTAGE WAS LOWER THAN EXPECTED, BUT STILL SUPPORTED FULL DEVICE FUNCTION. A SERIES OF AUTOMATED ELECTRICAL/FUNCTIONAL TESTS WERE CONDUCTED AND NO ISSUES WITH DEVICE PERFORMANCE WERE OBSERVED; BASIC SENSING, PACING AND SHOCKING FUNCTIONS OF THE DEVICE WERE VERIFIED. USING HISTORICAL DAILY BATTERY VOLTAGE MEASUREMENT DATA, ENGINEERS DETERMINED THAT THIS DEVICE WAS DEMONSTRATING BEHAVIOR CONSISTENT WITH A HIGH CURRENT CONDITION ASSOCIATED WITH A COMPROMISED LOW VOLTAGE CAPACITOR CONNECTED TO THE DEVICE'S BATTERY. LOW VOLTAGE CAPACITORS ARE USED IN THE DEVICE'S HIGH VOLTAGE CHARGING OPERATION IN ORDER TO FACILITATE FAST CHARGE TIMES. MALFUNCTION OF THESE CAPACITORS RESULTED IN A HIGH CURRENT DRAIN, WHICH WAS DEPLETING THIS DEVICE'S BATTERY FASTER THAN NORMAL.

Additional Manufacturer Narrative · 1

(B)(4). THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE RECORDED A CODE 1003 INDICATIVE OF BATTERY VOLTAGE TOO LOW FOR THE PROJECTED REMAINING CAPACITY. DATA ANALYSIS SHOWED THE BATTERY APPEARED TO BE DEPLETING MORE QUICKLY THAN EXPECTED. DEVICE REPLACEMENT WAS RECOMMENDED. THIS DEVICE REMAINS IN SERVICE. THERE HAVE BEEN NO ADVERSE PATIENT EFFECTS. BOSTON SCIENTIFIC HAS ISSUED AN ADVISORY COMMUNICATION REGARDING AN OLDER SUBSET OF COGNIS/TELIGEN DEVICES THAT IS MORE SUSCEPTIBLE TO A PRODUCT PERFORMANCE ANOMALY. SPECIFICALLY, THE PERFORMANCE OF A LOW VOLTAGE CAPACITOR MAY BE COMPROMISED OVER TIME, CAUSING AN INCREASED CURRENT DRAIN THAT CAN LEAD TO PREMATURE BATTERY DEPLETION. THIS PARTICULAR DEVICE WAS INCLUDED IN THE ADVISORY POPULATION.

Description of Event or Problem · 1

INFORMATION INDICATES THIS DEVICE HAS BEEN EXPLANTED. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
427453 COGNIS IMPLANTABLE CHF PULSE GENERATOR LWP GUIDANT CRM CLONMEL IRELAND N119

Patients

Seq Age Sex Outcome Treatment
1 71 YR N119| 7120| 4194| 1882TC