FDA Adverse Event Injury Summary report: N

INGENIO

MDR report key: 4884736 · Received July 1, 2015

Report

Report Number
2124215-2015-07371
Event Type
Injury
Date Received
July 1, 2015
Date of Event
May 27, 2015
Report Date
December 30, 2017
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. ANALYSIS DID NOT IDENTIFY ANY DEVICE CHARACTERISTICS THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED CLINICAL OBSERVATIONS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PATIENTS PACEMAKER AND RIGHT ATRIAL (RA) LEAD EXHIBITED OVERSENSING AND NOISE, ALONG WITH HIGH OUT-OF-RANGE (OOR) PACING IMPEDANCES OF GREATER THAN 2000 OHMS. AS A RESULT THE LEAD WAS SURGICALLY ABANDONED AND SUCCESSFULLY REPLACED. TO DATE, NO ADDITIONAL ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED.

Description of Event or Problem · 1

APPROXIMATELY TWO YEARS LATER THIS DEVICE WAS EXPLANTED AND RETURNED FOR ANALYSIS. NO ADDITIONAL INFORMATION WAS RECEIVED REGARDING THE REASON FOR EXPLANT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
428162 INGENIO IMPLANTABLE PULSE GENERATOR LWP GUIDANT CRM CLONMEL IRELAND K173

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization| L| R (B)(4)| (B)(4)| (B)(4)