TELIGEN
Report
- Report Number
- 2124215-2015-07103
- Event Type
- Injury
- Date Received
- July 1, 2015
- Date of Event
- May 20, 2015
- Report Date
- June 5, 2015
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, VISUAL INSPECTION OF THE ICD IDENTIFIED NO ANOMALIES. REVIEW OF DEVICE MEMORY FOUND A SHORTED LEAD FAULT (FAULT CODE 1005) WAS RECORDED. AN X-RAY OF THE DEVICE REVEALED THAT THE INTERNAL HIGH VOLTAGE FUSE WAS DAMAGED (I.E., NO LONGER INTACT). THE DAMAGE TO THE FUSE MOST LIKELY OCCURRED DURING DELIVERY OF THE SHOCK THAT RESULTED IN THE SHORTED SHOCK LEAD FAULT. THE DAMAGE TO THE HIGH VOLTAGE FUSE PREVENTED ANY SUBSEQUENT CHARGING OF THE HIGH VOLTAGE CAPACITORS AND DELIVERY OF SHOCK THERAPY.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT IMPLANTED WITH THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND RIGHT VENTRICULAR (RV) LEAD COMPLAINED OF CHEST PAIN. UPON INTERROGATION OF THE ICD WITH A PROGRAMMER, A SHOCK LEAD SHORTED MESSAGE WAS OBSERVED. REVIEW OF STORED DEVICE MEMORY REVEALED LOW OUT OF RANGE SHOCK IMPEDANCE MEASUREMENTS LESS THAN 20 OHMS DURING DELIVERY OF APPROPRIATE ANTI-TACHYCARDIA PACING (ATP) AND SHOCK THERAPY. AN INVASIVE PROCEDURE WAS PERFORMED. THE DEVICE WAS EXPLANTED AND REPLACED AND THE LEAD WAS SURGICALLY ABANDONED AND REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 426222 | TELIGEN | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | E110 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR | Hospitalization| L| R | 0175| E110| 4135 |